A triple-head SPECT system with parallel-hole collimators of different acceptance angles

Zongjian Cao, Lin Qian

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

We proposed to use three different parallel-hole collimators for a triple-head SPECT system. One of the collimators had a small collimator acceptance angle to provide ultra-high spatial resolution and the other two had larger collimator acceptance angles to achieve high counts. A new 2D reconstruction algorithm that combined the data acquired from different collimators was derived to take advantages of both high resolution and high sensitivity. The algorithm was evaluated using a computer-simulated matrix of spherical sources. For noise-free data, the accuracy (mainly determined by spatial resolution) obtained from combination of the collimator acceptance angles of 1.35, 4.05 and 6.75°(or 1.35, 5.40 and 9.45°) was slightly inferior to that obtained from three same LEHR collimators (with a 2.70°collimator acceptance angle). This is because the modulation transfer function (MTF) resulting from three different collimators decreases more quickly at low frequencies but becomes comparable at high frequencies as compared with the MTF of the 2.70°collimator. With noisy data, however, the image quality obtained with three different collimators was better than that resulting from any combinations of three same collimators. The improvement was only achieved by using the derived algorithm, while the conventional FBP algorithm did not improve image quality even with the same collimator configuration.

Original languageEnglish (US)
Pages (from-to)145-157
Number of pages13
JournalInternational Journal of Medical Informatics
Volume46
Issue number3
DOIs
StatePublished - Oct 1 1997

Fingerprint

Single-Photon Emission-Computed Tomography
Head
Noise

Keywords

  • Acceptance angles
  • Parallel-hole collimators
  • Spect system

ASJC Scopus subject areas

  • Medicine(all)

Cite this

A triple-head SPECT system with parallel-hole collimators of different acceptance angles. / Cao, Zongjian; Qian, Lin.

In: International Journal of Medical Informatics, Vol. 46, No. 3, 01.10.1997, p. 145-157.

Research output: Contribution to journalArticle

@article{3e2e3727b7e54c9cb4e776e3d531412d,
title = "A triple-head SPECT system with parallel-hole collimators of different acceptance angles",
abstract = "We proposed to use three different parallel-hole collimators for a triple-head SPECT system. One of the collimators had a small collimator acceptance angle to provide ultra-high spatial resolution and the other two had larger collimator acceptance angles to achieve high counts. A new 2D reconstruction algorithm that combined the data acquired from different collimators was derived to take advantages of both high resolution and high sensitivity. The algorithm was evaluated using a computer-simulated matrix of spherical sources. For noise-free data, the accuracy (mainly determined by spatial resolution) obtained from combination of the collimator acceptance angles of 1.35, 4.05 and 6.75°(or 1.35, 5.40 and 9.45°) was slightly inferior to that obtained from three same LEHR collimators (with a 2.70°collimator acceptance angle). This is because the modulation transfer function (MTF) resulting from three different collimators decreases more quickly at low frequencies but becomes comparable at high frequencies as compared with the MTF of the 2.70°collimator. With noisy data, however, the image quality obtained with three different collimators was better than that resulting from any combinations of three same collimators. The improvement was only achieved by using the derived algorithm, while the conventional FBP algorithm did not improve image quality even with the same collimator configuration.",
keywords = "Acceptance angles, Parallel-hole collimators, Spect system",
author = "Zongjian Cao and Lin Qian",
year = "1997",
month = "10",
day = "1",
doi = "10.1016/S1386-5056(97)00067-1",
language = "English (US)",
volume = "46",
pages = "145--157",
journal = "International Journal of Medical Informatics",
issn = "1386-5056",
publisher = "Elsevier Ireland Ltd",
number = "3",

}

TY - JOUR

T1 - A triple-head SPECT system with parallel-hole collimators of different acceptance angles

AU - Cao, Zongjian

AU - Qian, Lin

PY - 1997/10/1

Y1 - 1997/10/1

N2 - We proposed to use three different parallel-hole collimators for a triple-head SPECT system. One of the collimators had a small collimator acceptance angle to provide ultra-high spatial resolution and the other two had larger collimator acceptance angles to achieve high counts. A new 2D reconstruction algorithm that combined the data acquired from different collimators was derived to take advantages of both high resolution and high sensitivity. The algorithm was evaluated using a computer-simulated matrix of spherical sources. For noise-free data, the accuracy (mainly determined by spatial resolution) obtained from combination of the collimator acceptance angles of 1.35, 4.05 and 6.75°(or 1.35, 5.40 and 9.45°) was slightly inferior to that obtained from three same LEHR collimators (with a 2.70°collimator acceptance angle). This is because the modulation transfer function (MTF) resulting from three different collimators decreases more quickly at low frequencies but becomes comparable at high frequencies as compared with the MTF of the 2.70°collimator. With noisy data, however, the image quality obtained with three different collimators was better than that resulting from any combinations of three same collimators. The improvement was only achieved by using the derived algorithm, while the conventional FBP algorithm did not improve image quality even with the same collimator configuration.

AB - We proposed to use three different parallel-hole collimators for a triple-head SPECT system. One of the collimators had a small collimator acceptance angle to provide ultra-high spatial resolution and the other two had larger collimator acceptance angles to achieve high counts. A new 2D reconstruction algorithm that combined the data acquired from different collimators was derived to take advantages of both high resolution and high sensitivity. The algorithm was evaluated using a computer-simulated matrix of spherical sources. For noise-free data, the accuracy (mainly determined by spatial resolution) obtained from combination of the collimator acceptance angles of 1.35, 4.05 and 6.75°(or 1.35, 5.40 and 9.45°) was slightly inferior to that obtained from three same LEHR collimators (with a 2.70°collimator acceptance angle). This is because the modulation transfer function (MTF) resulting from three different collimators decreases more quickly at low frequencies but becomes comparable at high frequencies as compared with the MTF of the 2.70°collimator. With noisy data, however, the image quality obtained with three different collimators was better than that resulting from any combinations of three same collimators. The improvement was only achieved by using the derived algorithm, while the conventional FBP algorithm did not improve image quality even with the same collimator configuration.

KW - Acceptance angles

KW - Parallel-hole collimators

KW - Spect system

UR - http://www.scopus.com/inward/record.url?scp=0031259589&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031259589&partnerID=8YFLogxK

U2 - 10.1016/S1386-5056(97)00067-1

DO - 10.1016/S1386-5056(97)00067-1

M3 - Article

C2 - 9373777

AN - SCOPUS:0031259589

VL - 46

SP - 145

EP - 157

JO - International Journal of Medical Informatics

JF - International Journal of Medical Informatics

SN - 1386-5056

IS - 3

ER -