An analytical reconstruction algorithm for multifocal converging-beam SPECT

Zongjian Cao, B. M W Tsui

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

Multifocal converging-beam collimation has been suggested for cardiac SPECT imaging to increase sensitivity over the heart without truncation of the activity distribution in the chest. In this study, an analytical reconstruction algorithm is derived for multifocal fan-beam and multifocal cone-beam tomography. In the algorithm, the projection data are differently weighted and filtered, depending on the distance from the detector. For a given image point, the set of filtered data corresponding to the distance between this point and detector is backprojected to determine the pixel value of the point. Thus, the backprojection is done only once at each projection view. To evaluate the algorithm, simulation studies are performed using a 3D Defrise slab phantom without considering photon attenuation and scatter, detector response, and statistical noise. Reconstructed images demonstrate that reasonable quality can be achieved with a modest focal-length variation rate and with a small radius of rotation. A collimator with a focal length increasing quickly near its centre provides better quality in the image region distant from the central plane of the cone geometry, but produces more severe artifacts at the centre of the reconstructed image, compared to a collimator with an initially slowly varying focal length.

Original languageEnglish (US)
Article number005
Pages (from-to)281-291
Number of pages11
JournalPhysics in Medicine and Biology
Volume39
Issue number2
DOIs
StatePublished - Dec 1 1994

Fingerprint

Single-Photon Emission-Computed Tomography
Detectors
Cones
collimators
detectors
cones
projection
Photons
Artifacts
Fans
Tomography
chest
collimation
Thorax
Pixels
fans
Imaging techniques
artifacts
slabs
Geometry

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Physics and Astronomy (miscellaneous)
  • Biomedical Engineering

Cite this

An analytical reconstruction algorithm for multifocal converging-beam SPECT. / Cao, Zongjian; Tsui, B. M W.

In: Physics in Medicine and Biology, Vol. 39, No. 2, 005, 01.12.1994, p. 281-291.

Research output: Contribution to journalArticle

Cao, Zongjian ; Tsui, B. M W. / An analytical reconstruction algorithm for multifocal converging-beam SPECT. In: Physics in Medicine and Biology. 1994 ; Vol. 39, No. 2. pp. 281-291.
@article{962f6d6fd1c24e389d186efe37bfdc9d,
title = "An analytical reconstruction algorithm for multifocal converging-beam SPECT",
abstract = "Multifocal converging-beam collimation has been suggested for cardiac SPECT imaging to increase sensitivity over the heart without truncation of the activity distribution in the chest. In this study, an analytical reconstruction algorithm is derived for multifocal fan-beam and multifocal cone-beam tomography. In the algorithm, the projection data are differently weighted and filtered, depending on the distance from the detector. For a given image point, the set of filtered data corresponding to the distance between this point and detector is backprojected to determine the pixel value of the point. Thus, the backprojection is done only once at each projection view. To evaluate the algorithm, simulation studies are performed using a 3D Defrise slab phantom without considering photon attenuation and scatter, detector response, and statistical noise. Reconstructed images demonstrate that reasonable quality can be achieved with a modest focal-length variation rate and with a small radius of rotation. A collimator with a focal length increasing quickly near its centre provides better quality in the image region distant from the central plane of the cone geometry, but produces more severe artifacts at the centre of the reconstructed image, compared to a collimator with an initially slowly varying focal length.",
author = "Zongjian Cao and Tsui, {B. M W}",
year = "1994",
month = "12",
day = "1",
doi = "10.1088/0031-9155/39/2/005",
language = "English (US)",
volume = "39",
pages = "281--291",
journal = "Physics in Medicine and Biology",
issn = "0031-9155",
publisher = "IOP Publishing Ltd.",
number = "2",

}

TY - JOUR

T1 - An analytical reconstruction algorithm for multifocal converging-beam SPECT

AU - Cao, Zongjian

AU - Tsui, B. M W

PY - 1994/12/1

Y1 - 1994/12/1

N2 - Multifocal converging-beam collimation has been suggested for cardiac SPECT imaging to increase sensitivity over the heart without truncation of the activity distribution in the chest. In this study, an analytical reconstruction algorithm is derived for multifocal fan-beam and multifocal cone-beam tomography. In the algorithm, the projection data are differently weighted and filtered, depending on the distance from the detector. For a given image point, the set of filtered data corresponding to the distance between this point and detector is backprojected to determine the pixel value of the point. Thus, the backprojection is done only once at each projection view. To evaluate the algorithm, simulation studies are performed using a 3D Defrise slab phantom without considering photon attenuation and scatter, detector response, and statistical noise. Reconstructed images demonstrate that reasonable quality can be achieved with a modest focal-length variation rate and with a small radius of rotation. A collimator with a focal length increasing quickly near its centre provides better quality in the image region distant from the central plane of the cone geometry, but produces more severe artifacts at the centre of the reconstructed image, compared to a collimator with an initially slowly varying focal length.

AB - Multifocal converging-beam collimation has been suggested for cardiac SPECT imaging to increase sensitivity over the heart without truncation of the activity distribution in the chest. In this study, an analytical reconstruction algorithm is derived for multifocal fan-beam and multifocal cone-beam tomography. In the algorithm, the projection data are differently weighted and filtered, depending on the distance from the detector. For a given image point, the set of filtered data corresponding to the distance between this point and detector is backprojected to determine the pixel value of the point. Thus, the backprojection is done only once at each projection view. To evaluate the algorithm, simulation studies are performed using a 3D Defrise slab phantom without considering photon attenuation and scatter, detector response, and statistical noise. Reconstructed images demonstrate that reasonable quality can be achieved with a modest focal-length variation rate and with a small radius of rotation. A collimator with a focal length increasing quickly near its centre provides better quality in the image region distant from the central plane of the cone geometry, but produces more severe artifacts at the centre of the reconstructed image, compared to a collimator with an initially slowly varying focal length.

UR - http://www.scopus.com/inward/record.url?scp=0028295262&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0028295262&partnerID=8YFLogxK

U2 - 10.1088/0031-9155/39/2/005

DO - 10.1088/0031-9155/39/2/005

M3 - Article

VL - 39

SP - 281

EP - 291

JO - Physics in Medicine and Biology

JF - Physics in Medicine and Biology

SN - 0031-9155

IS - 2

M1 - 005

ER -