T1 - Clinical experience with a chlorhexidine-coated PICC
T2 - A prospective, multicenter, observational study
AU - Winkler, Michael A.
AU - Spencer, Timothy R.
AU - Siddiqi, Nasir
AU - Wallace, Jessica E.
AU - Gallien, John Z.
AU - Elbalasi, Hossam
AU - Issa, Mohamed
AU - Yu, Qian
AU - Raissi, Driss
N1 - Funding Information:
The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This study was sponsored by Teleflex Inc., (Morrisville, NC, USA), manufacturer of the chlorhexidine-coated PICC used in this study. The primary investigator institutions involved received a research grant to perform this study.
Publisher Copyright:
© The Author(s) 2021.
PY - 2021
Y1 - 2021
N2 - Introduction: This study evaluated a chlorhexidine-coated peripherally inserted central catheter (PICC) and the incidence of associated complications within both inpatient and outpatient populations. Methods: This IRB-approved, multicenter, prospective observational study was performed at three large teaching hospitals in the US. All adults who required a PICC for ⩾14 days were considered. Patients were monitored throughout entire catheter dwell. Duplex venous ultrasounds were performed before insertion, after 10 to 14 days of dwell time, and upon removal. Data was collected from the hospital, outpatient clinic, and patient PICC diary records. Results: A total of 103 patients, 56% male, with mean BMI 29 ± 8.8, were enrolled. The majority (79%) of patients were from high-risk groups—cancer, infectious diseases, transplant, and trauma. Primary treatment indications were antibiotics (66.99%) and chemotherapy (25.24%). Double lumen PICCs (59.2%) were favored clinically, as was basilic vein placement (71.84%). Mean catheter dwell was 47.01 ± 25.82 days. Three (3, 2.9%) central line-associated bloodstream infections (CLABSI) were reported. Four patients (4.6%) reported symptomatic catheter-related thrombosis (CRT), confirmed with ultrasound. Three patients (3.4%) had ultrasound-confirmed fibroblastic sleeve (FS). Eight patients (9.2%) who entered the study with pre-existing superficial thrombosis, had complete resolution at the time of catheter removal. The incidence of CLABSI was 0.82/1000 days. The combined CRT and FS rate was 6.9%. Conclusion: Based upon the observational findings of this study, chlorhexidine-coated PICC technology may be considered for use in patient populations who are at moderate to high-risk for catheter-related complications in both inpatient and outpatient settings.
AB - Introduction: This study evaluated a chlorhexidine-coated peripherally inserted central catheter (PICC) and the incidence of associated complications within both inpatient and outpatient populations. Methods: This IRB-approved, multicenter, prospective observational study was performed at three large teaching hospitals in the US. All adults who required a PICC for ⩾14 days were considered. Patients were monitored throughout entire catheter dwell. Duplex venous ultrasounds were performed before insertion, after 10 to 14 days of dwell time, and upon removal. Data was collected from the hospital, outpatient clinic, and patient PICC diary records. Results: A total of 103 patients, 56% male, with mean BMI 29 ± 8.8, were enrolled. The majority (79%) of patients were from high-risk groups—cancer, infectious diseases, transplant, and trauma. Primary treatment indications were antibiotics (66.99%) and chemotherapy (25.24%). Double lumen PICCs (59.2%) were favored clinically, as was basilic vein placement (71.84%). Mean catheter dwell was 47.01 ± 25.82 days. Three (3, 2.9%) central line-associated bloodstream infections (CLABSI) were reported. Four patients (4.6%) reported symptomatic catheter-related thrombosis (CRT), confirmed with ultrasound. Three patients (3.4%) had ultrasound-confirmed fibroblastic sleeve (FS). Eight patients (9.2%) who entered the study with pre-existing superficial thrombosis, had complete resolution at the time of catheter removal. The incidence of CLABSI was 0.82/1000 days. The combined CRT and FS rate was 6.9%. Conclusion: Based upon the observational findings of this study, chlorhexidine-coated PICC technology may be considered for use in patient populations who are at moderate to high-risk for catheter-related complications in both inpatient and outpatient settings.
KW - Peripherally inserted central catheter
KW - anti-thrombogenic
KW - antimicrobial
KW - colonization
KW - prevention
KW - thrombosis
UR - http://www.scopus.com/inward/record.url?scp=85116776205&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85116776205&partnerID=8YFLogxK
U2 - 10.1177/11297298211049648
DO - 10.1177/11297298211049648
M3 - Article
AN - SCOPUS:85116776205
SN - 1129-7298
JO - Journal of Vascular Access
JF - Journal of Vascular Access
ER -