Dose verification of stereotactic radiosurgery treatment for trigeminal neuralgia with presage 3D dosimetry system

Z. Wang, A. Thomas, Joseph R Newton, G. Ibbott, J. Deasy, M. Oldham

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Achieving adequate verification and quality-assurance (QA) for radiosurgery treatment of trigeminal-neuralgia (TGN) is particularly challenging because of the combination of very small fields, very high doses, and complex irradiation geometries (multiple gantry and couch combinations). TGN treatments have extreme requirements for dosimetry tools and QA techniques, to ensure adequate verification. In this work we evaluate the potential of Presage/Optical-CT dosimetry system as a tool for the verification of TGN distributions in high-resolution and in 3D. A TGN treatment was planned and delivered to a Presage 3D dosimeter positioned inside the Radiological-Physics-Center (RPC) head and neck IMRT credentialing phantom. A 6-arc treatment plan was created using the iPlan system, and a maximum dose of 80Gy was delivered with a Varian Trilogy machine. The delivered dose to Presage was determined by optical-CT scanning using the Duke Large field-of-view Optical-CT Scanner (DLOS) in 3D, with isotropic resolution of 0.7mm3. DLOS scanning and reconstruction took about 20minutes. 3D dose comparisons were made with the planning system. Good agreement was observed between the planned and measured 3D dose distributions, and this work provides strong support for the viability of Presage/Optical-CT as a highly useful new approach for verification of this complex technique.

Original languageEnglish (US)
Article number012058
Pages (from-to)283-286
Number of pages4
JournalJournal of Physics: Conference Series
Volume250
DOIs
StatePublished - Jan 1 2010

Fingerprint

dosimeters
dosage
assurance
scanners
field of view
couches
gantry cranes
scanning
viability
planning
arcs
requirements
physics
irradiation
high resolution
geometry

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Cite this

Dose verification of stereotactic radiosurgery treatment for trigeminal neuralgia with presage 3D dosimetry system. / Wang, Z.; Thomas, A.; Newton, Joseph R; Ibbott, G.; Deasy, J.; Oldham, M.

In: Journal of Physics: Conference Series, Vol. 250, 012058, 01.01.2010, p. 283-286.

Research output: Contribution to journalArticle

@article{c1225a3a6de94f6f97f3441bb89035f2,
title = "Dose verification of stereotactic radiosurgery treatment for trigeminal neuralgia with presage 3D dosimetry system",
abstract = "Achieving adequate verification and quality-assurance (QA) for radiosurgery treatment of trigeminal-neuralgia (TGN) is particularly challenging because of the combination of very small fields, very high doses, and complex irradiation geometries (multiple gantry and couch combinations). TGN treatments have extreme requirements for dosimetry tools and QA techniques, to ensure adequate verification. In this work we evaluate the potential of Presage/Optical-CT dosimetry system as a tool for the verification of TGN distributions in high-resolution and in 3D. A TGN treatment was planned and delivered to a Presage 3D dosimeter positioned inside the Radiological-Physics-Center (RPC) head and neck IMRT credentialing phantom. A 6-arc treatment plan was created using the iPlan system, and a maximum dose of 80Gy was delivered with a Varian Trilogy machine. The delivered dose to Presage was determined by optical-CT scanning using the Duke Large field-of-view Optical-CT Scanner (DLOS) in 3D, with isotropic resolution of 0.7mm3. DLOS scanning and reconstruction took about 20minutes. 3D dose comparisons were made with the planning system. Good agreement was observed between the planned and measured 3D dose distributions, and this work provides strong support for the viability of Presage/Optical-CT as a highly useful new approach for verification of this complex technique.",
author = "Z. Wang and A. Thomas and Newton, {Joseph R} and G. Ibbott and J. Deasy and M. Oldham",
year = "2010",
month = "1",
day = "1",
doi = "10.1088/1742-6596/250/1/012058",
language = "English (US)",
volume = "250",
pages = "283--286",
journal = "Journal of Physics: Conference Series",
issn = "1742-6588",
publisher = "IOP Publishing Ltd.",

}

TY - JOUR

T1 - Dose verification of stereotactic radiosurgery treatment for trigeminal neuralgia with presage 3D dosimetry system

AU - Wang, Z.

AU - Thomas, A.

AU - Newton, Joseph R

AU - Ibbott, G.

AU - Deasy, J.

AU - Oldham, M.

PY - 2010/1/1

Y1 - 2010/1/1

N2 - Achieving adequate verification and quality-assurance (QA) for radiosurgery treatment of trigeminal-neuralgia (TGN) is particularly challenging because of the combination of very small fields, very high doses, and complex irradiation geometries (multiple gantry and couch combinations). TGN treatments have extreme requirements for dosimetry tools and QA techniques, to ensure adequate verification. In this work we evaluate the potential of Presage/Optical-CT dosimetry system as a tool for the verification of TGN distributions in high-resolution and in 3D. A TGN treatment was planned and delivered to a Presage 3D dosimeter positioned inside the Radiological-Physics-Center (RPC) head and neck IMRT credentialing phantom. A 6-arc treatment plan was created using the iPlan system, and a maximum dose of 80Gy was delivered with a Varian Trilogy machine. The delivered dose to Presage was determined by optical-CT scanning using the Duke Large field-of-view Optical-CT Scanner (DLOS) in 3D, with isotropic resolution of 0.7mm3. DLOS scanning and reconstruction took about 20minutes. 3D dose comparisons were made with the planning system. Good agreement was observed between the planned and measured 3D dose distributions, and this work provides strong support for the viability of Presage/Optical-CT as a highly useful new approach for verification of this complex technique.

AB - Achieving adequate verification and quality-assurance (QA) for radiosurgery treatment of trigeminal-neuralgia (TGN) is particularly challenging because of the combination of very small fields, very high doses, and complex irradiation geometries (multiple gantry and couch combinations). TGN treatments have extreme requirements for dosimetry tools and QA techniques, to ensure adequate verification. In this work we evaluate the potential of Presage/Optical-CT dosimetry system as a tool for the verification of TGN distributions in high-resolution and in 3D. A TGN treatment was planned and delivered to a Presage 3D dosimeter positioned inside the Radiological-Physics-Center (RPC) head and neck IMRT credentialing phantom. A 6-arc treatment plan was created using the iPlan system, and a maximum dose of 80Gy was delivered with a Varian Trilogy machine. The delivered dose to Presage was determined by optical-CT scanning using the Duke Large field-of-view Optical-CT Scanner (DLOS) in 3D, with isotropic resolution of 0.7mm3. DLOS scanning and reconstruction took about 20minutes. 3D dose comparisons were made with the planning system. Good agreement was observed between the planned and measured 3D dose distributions, and this work provides strong support for the viability of Presage/Optical-CT as a highly useful new approach for verification of this complex technique.

UR - http://www.scopus.com/inward/record.url?scp=78650561651&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=78650561651&partnerID=8YFLogxK

U2 - 10.1088/1742-6596/250/1/012058

DO - 10.1088/1742-6596/250/1/012058

M3 - Article

AN - SCOPUS:78650561651

VL - 250

SP - 283

EP - 286

JO - Journal of Physics: Conference Series

JF - Journal of Physics: Conference Series

SN - 1742-6588

M1 - 012058

ER -