Effects of photon attenuation on increased counts in fanbeam collimation

Zong Jian Cao, Nellie Kelty

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

This study examined the effects of photon attenuation on the increased sensitivity (gain of counts) produced by a fanbeam collimator with imaging geometry similar to clinical practice. Methods: The number of counts was measured in experiments and by computer simulations for two cylindrical phantoms of different sizes using a fanbeam and a parallel-hole (parallel-beam) collimator with the same spatial resolution. The phantoms were located either in air (simulating the brain) or in a cubic water tank (simulating a source in the body). The ratio of the counts from the fanbeam collimator versus the parallel-beam collimator was obtained as a function of the distance between the phantom and detector. Results: Even with attenuation inside the phantom, the count ratio for the phantom located in air remains almost the same as that theoretically predicted without attenuation. But the count ratio obtained in the cubic water tank was decreased compared to the predicted values. The amount of decrease (reduced sensitivity) depended on the distance from the detector, the size of the source and the depth of the source. Conclusion: For brain studies, attenuation caused little change in the gain of counts provided by a fanbeam collimator. When imaging a structure located deep within the body, however, the gain of counts (sensitivity) may be appreciably reduced due to a longer path of photons within the body which results in more photon attenuation in fanbeam geometry.

Original languageEnglish (US)
Pages (from-to)187-190
Number of pages4
JournalJournal of Nuclear Medicine Technology
Volume25
Issue number3
StatePublished - Sep 1 1997

Fingerprint

Photons
Air
Water
Brain
Computer Simulation

Keywords

  • Attenuation
  • Fanbeam collimation
  • Sensitivity
  • SPECT

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging

Cite this

Effects of photon attenuation on increased counts in fanbeam collimation. / Cao, Zong Jian; Kelty, Nellie.

In: Journal of Nuclear Medicine Technology, Vol. 25, No. 3, 01.09.1997, p. 187-190.

Research output: Contribution to journalArticle

Cao, Zong Jian ; Kelty, Nellie. / Effects of photon attenuation on increased counts in fanbeam collimation. In: Journal of Nuclear Medicine Technology. 1997 ; Vol. 25, No. 3. pp. 187-190.
@article{64833ba429bb40b3a6b7c99057868c1a,
title = "Effects of photon attenuation on increased counts in fanbeam collimation",
abstract = "This study examined the effects of photon attenuation on the increased sensitivity (gain of counts) produced by a fanbeam collimator with imaging geometry similar to clinical practice. Methods: The number of counts was measured in experiments and by computer simulations for two cylindrical phantoms of different sizes using a fanbeam and a parallel-hole (parallel-beam) collimator with the same spatial resolution. The phantoms were located either in air (simulating the brain) or in a cubic water tank (simulating a source in the body). The ratio of the counts from the fanbeam collimator versus the parallel-beam collimator was obtained as a function of the distance between the phantom and detector. Results: Even with attenuation inside the phantom, the count ratio for the phantom located in air remains almost the same as that theoretically predicted without attenuation. But the count ratio obtained in the cubic water tank was decreased compared to the predicted values. The amount of decrease (reduced sensitivity) depended on the distance from the detector, the size of the source and the depth of the source. Conclusion: For brain studies, attenuation caused little change in the gain of counts provided by a fanbeam collimator. When imaging a structure located deep within the body, however, the gain of counts (sensitivity) may be appreciably reduced due to a longer path of photons within the body which results in more photon attenuation in fanbeam geometry.",
keywords = "Attenuation, Fanbeam collimation, Sensitivity, SPECT",
author = "Cao, {Zong Jian} and Nellie Kelty",
year = "1997",
month = "9",
day = "1",
language = "English (US)",
volume = "25",
pages = "187--190",
journal = "Journal of Nuclear Medicine Technology",
issn = "0091-4916",
publisher = "Society of Nuclear Medicine Inc.",
number = "3",

}

TY - JOUR

T1 - Effects of photon attenuation on increased counts in fanbeam collimation

AU - Cao, Zong Jian

AU - Kelty, Nellie

PY - 1997/9/1

Y1 - 1997/9/1

N2 - This study examined the effects of photon attenuation on the increased sensitivity (gain of counts) produced by a fanbeam collimator with imaging geometry similar to clinical practice. Methods: The number of counts was measured in experiments and by computer simulations for two cylindrical phantoms of different sizes using a fanbeam and a parallel-hole (parallel-beam) collimator with the same spatial resolution. The phantoms were located either in air (simulating the brain) or in a cubic water tank (simulating a source in the body). The ratio of the counts from the fanbeam collimator versus the parallel-beam collimator was obtained as a function of the distance between the phantom and detector. Results: Even with attenuation inside the phantom, the count ratio for the phantom located in air remains almost the same as that theoretically predicted without attenuation. But the count ratio obtained in the cubic water tank was decreased compared to the predicted values. The amount of decrease (reduced sensitivity) depended on the distance from the detector, the size of the source and the depth of the source. Conclusion: For brain studies, attenuation caused little change in the gain of counts provided by a fanbeam collimator. When imaging a structure located deep within the body, however, the gain of counts (sensitivity) may be appreciably reduced due to a longer path of photons within the body which results in more photon attenuation in fanbeam geometry.

AB - This study examined the effects of photon attenuation on the increased sensitivity (gain of counts) produced by a fanbeam collimator with imaging geometry similar to clinical practice. Methods: The number of counts was measured in experiments and by computer simulations for two cylindrical phantoms of different sizes using a fanbeam and a parallel-hole (parallel-beam) collimator with the same spatial resolution. The phantoms were located either in air (simulating the brain) or in a cubic water tank (simulating a source in the body). The ratio of the counts from the fanbeam collimator versus the parallel-beam collimator was obtained as a function of the distance between the phantom and detector. Results: Even with attenuation inside the phantom, the count ratio for the phantom located in air remains almost the same as that theoretically predicted without attenuation. But the count ratio obtained in the cubic water tank was decreased compared to the predicted values. The amount of decrease (reduced sensitivity) depended on the distance from the detector, the size of the source and the depth of the source. Conclusion: For brain studies, attenuation caused little change in the gain of counts provided by a fanbeam collimator. When imaging a structure located deep within the body, however, the gain of counts (sensitivity) may be appreciably reduced due to a longer path of photons within the body which results in more photon attenuation in fanbeam geometry.

KW - Attenuation

KW - Fanbeam collimation

KW - Sensitivity

KW - SPECT

UR - http://www.scopus.com/inward/record.url?scp=0030765607&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030765607&partnerID=8YFLogxK

M3 - Article

C2 - 9291064

AN - SCOPUS:0030765607

VL - 25

SP - 187

EP - 190

JO - Journal of Nuclear Medicine Technology

JF - Journal of Nuclear Medicine Technology

SN - 0091-4916

IS - 3

ER -