N2 - The purpose of this study was to test the feasibility of adapting a new microtensile testing technique to measure resin cement bond strengths to the cervical, middle, and apical thirds of root canals. Post spaces were created in extracted human teeth, and the roots were ground flat on one side to expose the canal and permit ideal placement of one of two resin cements (Panavia 21 or C&B Metabond). After 48 h of storage, serial 1-mm-thick cross-sections were cut to create 6-10 specimens per root. The first three specimens were from the cervical third, the next three were from the middle third, and the last three were from the apical third of the root. Each 1 × 1 × 8 mm specimen was pulled to failure in a miniature testing machine. The results indicated that both resin cements produced high bond strengths (12-23 MPa), and that bond strengths to the apical third were significantly higher (p < 0.05) than to the cervical or middle third with either cement. This new method shows promise for evaluating resin bond strengths within root canals.
AB - The purpose of this study was to test the feasibility of adapting a new microtensile testing technique to measure resin cement bond strengths to the cervical, middle, and apical thirds of root canals. Post spaces were created in extracted human teeth, and the roots were ground flat on one side to expose the canal and permit ideal placement of one of two resin cements (Panavia 21 or C&B Metabond). After 48 h of storage, serial 1-mm-thick cross-sections were cut to create 6-10 specimens per root. The first three specimens were from the cervical third, the next three were from the middle third, and the last three were from the apical third of the root. Each 1 × 1 × 8 mm specimen was pulled to failure in a miniature testing machine. The results indicated that both resin cements produced high bond strengths (12-23 MPa), and that bond strengths to the apical third were significantly higher (p < 0.05) than to the cervical or middle third with either cement. This new method shows promise for evaluating resin bond strengths within root canals.
UR - http://www.scopus.com/inward/record.url?scp=0035346177&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0035346177&partnerID=8YFLogxK
U2 - 10.1097/00004770-200105000-00001
DO - 10.1097/00004770-200105000-00001
M3 - Article
C2 - 11485248
AN - SCOPUS:0035346177
SN - 0099-2399
VL - 27
SP - 321
EP - 324
JO - Journal of endodontics
JF - Journal of endodontics
IS - 5
ER -