Foil electret transducer for blood pressure monitoring

J. E. West, I. J. Busch-Vishniac, G. A. Marshfield, T. G. Pickerina

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

A foil electret microphone for use under the cuff of an automatic blood pressure monitoring - system is described. The transducer is designed to operate with relatively flat sensitivity over a static pressure range of 40 to 250 mm Hg (5.33 X104to 3.33 X 10sdyn/cm2). The new electret microphone differs from conventional microphones used for airborne sound reception in two ways: (1) the diaphragm thickness is 50 fitn rather than the typical 12.5 or 25 fim, and (2) the backplate contains a set of annular ridges spaced at 4 mm rather than the typical 7–10 mm. This microphone offers three advantages over the piezoelectric microphone now in use: (1) greater tolerance'in positioning the microphone over the brachial artery, (2) nearly 20-dB higher sensitivity and signal-to-noise ratio, and (3) the ability to obtain measurements with the. microphone placed midway between the elbow and shoulder, Tests of the new foil electret microphone in conjunction with the automatic blood pressure monitoring system indicate that the automatic and conventional measurements of systolic and diastolic blood pressure agree to within 5 mm Hg at least 90% of the time. In addition, the electret microphone is able to obtain automatic measurements on subjects with a wider range of ages and sizes.

Original languageEnglish (US)
Pages (from-to)680-686
Number of pages7
JournalJournal of the Acoustical Society of America
Volume74
Issue number3
DOIs
StatePublished - Jan 1 1983
Externally publishedYes

Fingerprint

blood pressure
electrets
microphones
foils
transducers
diastolic pressure
cuffs
Monitoring
Blood Pressure
static pressure
sensitivity
diaphragms
shoulders
arteries
positioning
Conventional
ridges
signal to noise ratios
acoustics

ASJC Scopus subject areas

  • Arts and Humanities (miscellaneous)
  • Acoustics and Ultrasonics

Cite this

Foil electret transducer for blood pressure monitoring. / West, J. E.; Busch-Vishniac, I. J.; Marshfield, G. A.; Pickerina, T. G.

In: Journal of the Acoustical Society of America, Vol. 74, No. 3, 01.01.1983, p. 680-686.

Research output: Contribution to journalArticle

West, J. E. ; Busch-Vishniac, I. J. ; Marshfield, G. A. ; Pickerina, T. G. / Foil electret transducer for blood pressure monitoring. In: Journal of the Acoustical Society of America. 1983 ; Vol. 74, No. 3. pp. 680-686.
@article{1195ab2602de4e8ea81f744a77108f78,
title = "Foil electret transducer for blood pressure monitoring",
abstract = "A foil electret microphone for use under the cuff of an automatic blood pressure monitoring - system is described. The transducer is designed to operate with relatively flat sensitivity over a static pressure range of 40 to 250 mm Hg (5.33 X104to 3.33 X 10sdyn/cm2). The new electret microphone differs from conventional microphones used for airborne sound reception in two ways: (1) the diaphragm thickness is 50 fitn rather than the typical 12.5 or 25 fim, and (2) the backplate contains a set of annular ridges spaced at 4 mm rather than the typical 7–10 mm. This microphone offers three advantages over the piezoelectric microphone now in use: (1) greater tolerance'in positioning the microphone over the brachial artery, (2) nearly 20-dB higher sensitivity and signal-to-noise ratio, and (3) the ability to obtain measurements with the. microphone placed midway between the elbow and shoulder, Tests of the new foil electret microphone in conjunction with the automatic blood pressure monitoring system indicate that the automatic and conventional measurements of systolic and diastolic blood pressure agree to within 5 mm Hg at least 90{\%} of the time. In addition, the electret microphone is able to obtain automatic measurements on subjects with a wider range of ages and sizes.",
author = "West, {J. E.} and Busch-Vishniac, {I. J.} and Marshfield, {G. A.} and Pickerina, {T. G.}",
year = "1983",
month = "1",
day = "1",
doi = "10.1121/1.389852",
language = "English (US)",
volume = "74",
pages = "680--686",
journal = "Journal of the Acoustical Society of America",
issn = "0001-4966",
publisher = "Acoustical Society of America",
number = "3",

}

TY - JOUR

T1 - Foil electret transducer for blood pressure monitoring

AU - West, J. E.

AU - Busch-Vishniac, I. J.

AU - Marshfield, G. A.

AU - Pickerina, T. G.

PY - 1983/1/1

Y1 - 1983/1/1

N2 - A foil electret microphone for use under the cuff of an automatic blood pressure monitoring - system is described. The transducer is designed to operate with relatively flat sensitivity over a static pressure range of 40 to 250 mm Hg (5.33 X104to 3.33 X 10sdyn/cm2). The new electret microphone differs from conventional microphones used for airborne sound reception in two ways: (1) the diaphragm thickness is 50 fitn rather than the typical 12.5 or 25 fim, and (2) the backplate contains a set of annular ridges spaced at 4 mm rather than the typical 7–10 mm. This microphone offers three advantages over the piezoelectric microphone now in use: (1) greater tolerance'in positioning the microphone over the brachial artery, (2) nearly 20-dB higher sensitivity and signal-to-noise ratio, and (3) the ability to obtain measurements with the. microphone placed midway between the elbow and shoulder, Tests of the new foil electret microphone in conjunction with the automatic blood pressure monitoring system indicate that the automatic and conventional measurements of systolic and diastolic blood pressure agree to within 5 mm Hg at least 90% of the time. In addition, the electret microphone is able to obtain automatic measurements on subjects with a wider range of ages and sizes.

AB - A foil electret microphone for use under the cuff of an automatic blood pressure monitoring - system is described. The transducer is designed to operate with relatively flat sensitivity over a static pressure range of 40 to 250 mm Hg (5.33 X104to 3.33 X 10sdyn/cm2). The new electret microphone differs from conventional microphones used for airborne sound reception in two ways: (1) the diaphragm thickness is 50 fitn rather than the typical 12.5 or 25 fim, and (2) the backplate contains a set of annular ridges spaced at 4 mm rather than the typical 7–10 mm. This microphone offers three advantages over the piezoelectric microphone now in use: (1) greater tolerance'in positioning the microphone over the brachial artery, (2) nearly 20-dB higher sensitivity and signal-to-noise ratio, and (3) the ability to obtain measurements with the. microphone placed midway between the elbow and shoulder, Tests of the new foil electret microphone in conjunction with the automatic blood pressure monitoring system indicate that the automatic and conventional measurements of systolic and diastolic blood pressure agree to within 5 mm Hg at least 90% of the time. In addition, the electret microphone is able to obtain automatic measurements on subjects with a wider range of ages and sizes.

UR - http://www.scopus.com/inward/record.url?scp=0020816187&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020816187&partnerID=8YFLogxK

U2 - 10.1121/1.389852

DO - 10.1121/1.389852

M3 - Article

C2 - 6630722

AN - SCOPUS:0020816187

VL - 74

SP - 680

EP - 686

JO - Journal of the Acoustical Society of America

JF - Journal of the Acoustical Society of America

SN - 0001-4966

IS - 3

ER -