Hemiazygous coil placement for high-defibrillation thresholds in a patient with a right-sided implantable cardioverter defibrillator

Kent R Nilsson, Kevin P. Jackson

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

A 41-year-old man underwent implantation of a right-sided implantable cardioverter defibrillator after removal of an infected left-sided system. Defibrillation threshold (DFT) testing on the right-sided system failed to convert ventricular fibrillation at maximum device output (35 J) compared with a DFT of less than 15 J on the previous left-sided system. A single-coil lead was selectively placed into the hemiazygous vein, which courses leftward of the spine in a posterior-anterior projection, resulting in an improved shocking vector and reduction in DFTs to less than 25 J.

Original languageEnglish (US)
JournalPACE - Pacing and Clinical Electrophysiology
Volume35
Issue number1
DOIs
StatePublished - Jan 1 2012
Externally publishedYes

Fingerprint

Implantable Defibrillators
Ventricular Fibrillation
Veins
Spine
Equipment and Supplies
Lead

Keywords

  • defibrillation threshold
  • hemiazygous vein
  • implantable cardioverter defibrillator

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

@article{10fce53d5ce24a0da03d388f899c1fc4,
title = "Hemiazygous coil placement for high-defibrillation thresholds in a patient with a right-sided implantable cardioverter defibrillator",
abstract = "A 41-year-old man underwent implantation of a right-sided implantable cardioverter defibrillator after removal of an infected left-sided system. Defibrillation threshold (DFT) testing on the right-sided system failed to convert ventricular fibrillation at maximum device output (35 J) compared with a DFT of less than 15 J on the previous left-sided system. A single-coil lead was selectively placed into the hemiazygous vein, which courses leftward of the spine in a posterior-anterior projection, resulting in an improved shocking vector and reduction in DFTs to less than 25 J.",
keywords = "defibrillation threshold, hemiazygous vein, implantable cardioverter defibrillator",
author = "Nilsson, {Kent R} and Jackson, {Kevin P.}",
year = "2012",
month = "1",
day = "1",
doi = "10.1111/j.1540-8159.2010.02862.x",
language = "English (US)",
volume = "35",
journal = "PACE - Pacing and Clinical Electrophysiology",
issn = "0147-8389",
publisher = "Wiley-Blackwell",
number = "1",

}

TY - JOUR

T1 - Hemiazygous coil placement for high-defibrillation thresholds in a patient with a right-sided implantable cardioverter defibrillator

AU - Nilsson, Kent R

AU - Jackson, Kevin P.

PY - 2012/1/1

Y1 - 2012/1/1

N2 - A 41-year-old man underwent implantation of a right-sided implantable cardioverter defibrillator after removal of an infected left-sided system. Defibrillation threshold (DFT) testing on the right-sided system failed to convert ventricular fibrillation at maximum device output (35 J) compared with a DFT of less than 15 J on the previous left-sided system. A single-coil lead was selectively placed into the hemiazygous vein, which courses leftward of the spine in a posterior-anterior projection, resulting in an improved shocking vector and reduction in DFTs to less than 25 J.

AB - A 41-year-old man underwent implantation of a right-sided implantable cardioverter defibrillator after removal of an infected left-sided system. Defibrillation threshold (DFT) testing on the right-sided system failed to convert ventricular fibrillation at maximum device output (35 J) compared with a DFT of less than 15 J on the previous left-sided system. A single-coil lead was selectively placed into the hemiazygous vein, which courses leftward of the spine in a posterior-anterior projection, resulting in an improved shocking vector and reduction in DFTs to less than 25 J.

KW - defibrillation threshold

KW - hemiazygous vein

KW - implantable cardioverter defibrillator

UR - http://www.scopus.com/inward/record.url?scp=79251582366&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79251582366&partnerID=8YFLogxK

U2 - 10.1111/j.1540-8159.2010.02862.x

DO - 10.1111/j.1540-8159.2010.02862.x

M3 - Article

C2 - 20723090

AN - SCOPUS:79251582366

VL - 35

JO - PACE - Pacing and Clinical Electrophysiology

JF - PACE - Pacing and Clinical Electrophysiology

SN - 0147-8389

IS - 1

ER -