Abstract

A Silastic electrode positioner was introduced by the Advanced Bionics Corporation in 1999 and it was designed to achieve a perimodiolar position of the stimulating electrode. The positioner was voluntarily recalled in the United States in July 2002 due to an apparent higher risk of bacterial meningitis in patients in whom the electrode positioner had been placed. A detailed histopathologic study of the positioner in the human has not previously been published. The histopathologic findings in a 74-year-old woman who underwent bilateral cochlear implantation using the positioner are presented. Findings include a large track caused by the combined electrode and its positioner with considerable disruption of the basilar membrane and osseous spiral lamina. Although there was a fibrous sheath around the electrode and positioner at the cochleostomy in both ears, this fibrous sheath did not extend deeply into the cochlea except at the apical end of the electrode beyond the positioner. This resulted in a large fluid space around and between the positioner and electrode within the cochlea and presumably in fluid continuity with the cerebrospinal fluid space. Possible clinical implications are discussed.

Original languageEnglish (US)
Pages (from-to)223-227
Number of pages5
JournalAudiology and Neurotology
Volume18
Issue number4
DOIs
StatePublished - Jan 1 2013
Externally publishedYes

Fingerprint

Cochlear Implants
Electrodes
Cochlea
Spiral Lamina
Bionics
Basilar Membrane
Cochlear Implantation
Bacterial Meningitides
Ear
Cerebrospinal Fluid

Keywords

  • Clarion® cochlear implant
  • Histopathology
  • Human temporal bone histopathology
  • Positioner

ASJC Scopus subject areas

  • Physiology
  • Otorhinolaryngology
  • Sensory Systems
  • Speech and Hearing

Cite this

Histopathology of the clarion cochlear implant electrode positioner in a human subject. / Seyyedi, Mohammad; Burgess, B. J.; Eddington, D. K.; Gantz, B. J.; Nadol, J. B.

In: Audiology and Neurotology, Vol. 18, No. 4, 01.01.2013, p. 223-227.

Research output: Contribution to journalArticle

Seyyedi, Mohammad ; Burgess, B. J. ; Eddington, D. K. ; Gantz, B. J. ; Nadol, J. B. / Histopathology of the clarion cochlear implant electrode positioner in a human subject. In: Audiology and Neurotology. 2013 ; Vol. 18, No. 4. pp. 223-227.
@article{dc723fd3155048f48da67b4212d73f15,
title = "Histopathology of the clarion cochlear implant electrode positioner in a human subject",
abstract = "A Silastic electrode positioner was introduced by the Advanced Bionics Corporation in 1999 and it was designed to achieve a perimodiolar position of the stimulating electrode. The positioner was voluntarily recalled in the United States in July 2002 due to an apparent higher risk of bacterial meningitis in patients in whom the electrode positioner had been placed. A detailed histopathologic study of the positioner in the human has not previously been published. The histopathologic findings in a 74-year-old woman who underwent bilateral cochlear implantation using the positioner are presented. Findings include a large track caused by the combined electrode and its positioner with considerable disruption of the basilar membrane and osseous spiral lamina. Although there was a fibrous sheath around the electrode and positioner at the cochleostomy in both ears, this fibrous sheath did not extend deeply into the cochlea except at the apical end of the electrode beyond the positioner. This resulted in a large fluid space around and between the positioner and electrode within the cochlea and presumably in fluid continuity with the cerebrospinal fluid space. Possible clinical implications are discussed.",
keywords = "Clarion{\circledR} cochlear implant, Histopathology, Human temporal bone histopathology, Positioner",
author = "Mohammad Seyyedi and Burgess, {B. J.} and Eddington, {D. K.} and Gantz, {B. J.} and Nadol, {J. B.}",
year = "2013",
month = "1",
day = "1",
doi = "10.1159/000351300",
language = "English (US)",
volume = "18",
pages = "223--227",
journal = "Audiology and Neuro-Otology",
issn = "1420-3030",
publisher = "S. Karger AG",
number = "4",

}

TY - JOUR

T1 - Histopathology of the clarion cochlear implant electrode positioner in a human subject

AU - Seyyedi, Mohammad

AU - Burgess, B. J.

AU - Eddington, D. K.

AU - Gantz, B. J.

AU - Nadol, J. B.

PY - 2013/1/1

Y1 - 2013/1/1

N2 - A Silastic electrode positioner was introduced by the Advanced Bionics Corporation in 1999 and it was designed to achieve a perimodiolar position of the stimulating electrode. The positioner was voluntarily recalled in the United States in July 2002 due to an apparent higher risk of bacterial meningitis in patients in whom the electrode positioner had been placed. A detailed histopathologic study of the positioner in the human has not previously been published. The histopathologic findings in a 74-year-old woman who underwent bilateral cochlear implantation using the positioner are presented. Findings include a large track caused by the combined electrode and its positioner with considerable disruption of the basilar membrane and osseous spiral lamina. Although there was a fibrous sheath around the electrode and positioner at the cochleostomy in both ears, this fibrous sheath did not extend deeply into the cochlea except at the apical end of the electrode beyond the positioner. This resulted in a large fluid space around and between the positioner and electrode within the cochlea and presumably in fluid continuity with the cerebrospinal fluid space. Possible clinical implications are discussed.

AB - A Silastic electrode positioner was introduced by the Advanced Bionics Corporation in 1999 and it was designed to achieve a perimodiolar position of the stimulating electrode. The positioner was voluntarily recalled in the United States in July 2002 due to an apparent higher risk of bacterial meningitis in patients in whom the electrode positioner had been placed. A detailed histopathologic study of the positioner in the human has not previously been published. The histopathologic findings in a 74-year-old woman who underwent bilateral cochlear implantation using the positioner are presented. Findings include a large track caused by the combined electrode and its positioner with considerable disruption of the basilar membrane and osseous spiral lamina. Although there was a fibrous sheath around the electrode and positioner at the cochleostomy in both ears, this fibrous sheath did not extend deeply into the cochlea except at the apical end of the electrode beyond the positioner. This resulted in a large fluid space around and between the positioner and electrode within the cochlea and presumably in fluid continuity with the cerebrospinal fluid space. Possible clinical implications are discussed.

KW - Clarion® cochlear implant

KW - Histopathology

KW - Human temporal bone histopathology

KW - Positioner

UR - http://www.scopus.com/inward/record.url?scp=84878903818&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84878903818&partnerID=8YFLogxK

U2 - 10.1159/000351300

DO - 10.1159/000351300

M3 - Article

VL - 18

SP - 223

EP - 227

JO - Audiology and Neuro-Otology

JF - Audiology and Neuro-Otology

SN - 1420-3030

IS - 4

ER -