Improved Image Quality for Asymmetric Double-focal Cone-beam SPECT

Zongjian Cao, B. M.W. Tsui

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

To optimize both spatial resolution and detection efficiency in brain SPECT imaging using a rectangular camera, an asymmetric double-focal cone-beam collimator is proposed with the focal points located near the base plane of the patient's head. To fit the entire head into the field-of-view of the collimator with dimensions of 50cmx40cm and at a radius-of-rotation of 15 cm, the focal lengths of the collimator are 55 and 70 cm, respectively, in the transverse and axial directions. With this geometry, the artifacts in the reconstructed image produced by the Feldkamp algorithm are more severe compared to those in a symmetric cone-beam geometry, due to the larger vertex angle between the top of the head and the base plane. To improve the reconstructed image quality, a fully three-dimensional (3D) reconstruction algorithm developed previously for single-focal cone-beam SPECT was extended to the asymmetric double-focal conebeam geometry. The algorithm involves nonstationary 2D filtering and a reprojection technique for estimation of the missing data caused by a single-orbit cone-beam geometry. The results from simulation studies with the 3D Defrise slab phantom demonstrated that the fully 3D algorithm provided a much improved image quality in terms of reduced slice-to-slice cross talks and shape elongation compared to that produced by the conventional Feldkamp algorithm.

Original languageEnglish (US)
Pages (from-to)1145-1148
Number of pages4
JournalIEEE Transactions on Nuclear Science
Volume40
Issue number4
DOIs
StatePublished - Jan 1 1993

Fingerprint

Image quality
Cones
cones
collimators
Geometry
geometry
elongation
field of view
brain
artifacts
Elongation
Brain
apexes
slabs
Orbits
spatial resolution
Cameras
cameras
orbits
Imaging techniques

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Nuclear Energy and Engineering
  • Electrical and Electronic Engineering

Cite this

Improved Image Quality for Asymmetric Double-focal Cone-beam SPECT. / Cao, Zongjian; Tsui, B. M.W.

In: IEEE Transactions on Nuclear Science, Vol. 40, No. 4, 01.01.1993, p. 1145-1148.

Research output: Contribution to journalArticle

Cao, Zongjian ; Tsui, B. M.W. / Improved Image Quality for Asymmetric Double-focal Cone-beam SPECT. In: IEEE Transactions on Nuclear Science. 1993 ; Vol. 40, No. 4. pp. 1145-1148.
@article{27e4be11f7c043c3b9dbb73ba8cc13da,
title = "Improved Image Quality for Asymmetric Double-focal Cone-beam SPECT",
abstract = "To optimize both spatial resolution and detection efficiency in brain SPECT imaging using a rectangular camera, an asymmetric double-focal cone-beam collimator is proposed with the focal points located near the base plane of the patient's head. To fit the entire head into the field-of-view of the collimator with dimensions of 50cmx40cm and at a radius-of-rotation of 15 cm, the focal lengths of the collimator are 55 and 70 cm, respectively, in the transverse and axial directions. With this geometry, the artifacts in the reconstructed image produced by the Feldkamp algorithm are more severe compared to those in a symmetric cone-beam geometry, due to the larger vertex angle between the top of the head and the base plane. To improve the reconstructed image quality, a fully three-dimensional (3D) reconstruction algorithm developed previously for single-focal cone-beam SPECT was extended to the asymmetric double-focal conebeam geometry. The algorithm involves nonstationary 2D filtering and a reprojection technique for estimation of the missing data caused by a single-orbit cone-beam geometry. The results from simulation studies with the 3D Defrise slab phantom demonstrated that the fully 3D algorithm provided a much improved image quality in terms of reduced slice-to-slice cross talks and shape elongation compared to that produced by the conventional Feldkamp algorithm.",
author = "Zongjian Cao and Tsui, {B. M.W.}",
year = "1993",
month = "1",
day = "1",
doi = "10.1109/23.256726",
language = "English (US)",
volume = "40",
pages = "1145--1148",
journal = "IEEE Transactions on Nuclear Science",
issn = "0018-9499",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "4",

}

TY - JOUR

T1 - Improved Image Quality for Asymmetric Double-focal Cone-beam SPECT

AU - Cao, Zongjian

AU - Tsui, B. M.W.

PY - 1993/1/1

Y1 - 1993/1/1

N2 - To optimize both spatial resolution and detection efficiency in brain SPECT imaging using a rectangular camera, an asymmetric double-focal cone-beam collimator is proposed with the focal points located near the base plane of the patient's head. To fit the entire head into the field-of-view of the collimator with dimensions of 50cmx40cm and at a radius-of-rotation of 15 cm, the focal lengths of the collimator are 55 and 70 cm, respectively, in the transverse and axial directions. With this geometry, the artifacts in the reconstructed image produced by the Feldkamp algorithm are more severe compared to those in a symmetric cone-beam geometry, due to the larger vertex angle between the top of the head and the base plane. To improve the reconstructed image quality, a fully three-dimensional (3D) reconstruction algorithm developed previously for single-focal cone-beam SPECT was extended to the asymmetric double-focal conebeam geometry. The algorithm involves nonstationary 2D filtering and a reprojection technique for estimation of the missing data caused by a single-orbit cone-beam geometry. The results from simulation studies with the 3D Defrise slab phantom demonstrated that the fully 3D algorithm provided a much improved image quality in terms of reduced slice-to-slice cross talks and shape elongation compared to that produced by the conventional Feldkamp algorithm.

AB - To optimize both spatial resolution and detection efficiency in brain SPECT imaging using a rectangular camera, an asymmetric double-focal cone-beam collimator is proposed with the focal points located near the base plane of the patient's head. To fit the entire head into the field-of-view of the collimator with dimensions of 50cmx40cm and at a radius-of-rotation of 15 cm, the focal lengths of the collimator are 55 and 70 cm, respectively, in the transverse and axial directions. With this geometry, the artifacts in the reconstructed image produced by the Feldkamp algorithm are more severe compared to those in a symmetric cone-beam geometry, due to the larger vertex angle between the top of the head and the base plane. To improve the reconstructed image quality, a fully three-dimensional (3D) reconstruction algorithm developed previously for single-focal cone-beam SPECT was extended to the asymmetric double-focal conebeam geometry. The algorithm involves nonstationary 2D filtering and a reprojection technique for estimation of the missing data caused by a single-orbit cone-beam geometry. The results from simulation studies with the 3D Defrise slab phantom demonstrated that the fully 3D algorithm provided a much improved image quality in terms of reduced slice-to-slice cross talks and shape elongation compared to that produced by the conventional Feldkamp algorithm.

UR - http://www.scopus.com/inward/record.url?scp=0027646317&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027646317&partnerID=8YFLogxK

U2 - 10.1109/23.256726

DO - 10.1109/23.256726

M3 - Article

VL - 40

SP - 1145

EP - 1148

JO - IEEE Transactions on Nuclear Science

JF - IEEE Transactions on Nuclear Science

SN - 0018-9499

IS - 4

ER -