Optimization of alarms

A study on alarm limits, alarm sounds, and false alarms, intended to reduce annoyance

Frank E Block, Lauri Nuutinen, Bert Ballast

Research output: Contribution to journalArticle

45 Citations (Scopus)

Abstract

Alarms in the operating room remain a major source of annoyance and confusion. Nearly all alarms result from a transgression of certain alarm limits. We surveyed manufacturers at a major meeting of anesthesiologists for their default alarm limits. We also conducted a mail survey of anesthesiologists in the United States, Finland, and the Netherlands, to learn their chosen alarm limits, reasons for turning off the alarms, and estimates of the number of false alarms. The surveys of medical monitoring manufacturers demonstrated a wide variety of default alarm settings. Anesthesiologists' chosen alarm limits tended to parallel the manufacturers' defaults. Fewer than 30% of anesthesiologists stated that they did not turn off the alarms. The leading reason for turning off alarms was the large number of false alarms. Estimates of the number of false alarms varied from 30% to 76%. The desired role of alarms in anesthetic practice and some suggestions for the general improvement of alarms are discussed.

Original languageEnglish (US)
Pages (from-to)75-83
Number of pages9
JournalJournal of Clinical Monitoring and Computing
Volume15
Issue number2
DOIs
StatePublished - Jun 26 1999
Externally publishedYes

Fingerprint

Postal Service
Operating Rooms
Finland
Netherlands
Anesthetics
Anesthesiologists
Surveys and Questionnaires

ASJC Scopus subject areas

  • Health Informatics
  • Critical Care and Intensive Care Medicine
  • Anesthesiology and Pain Medicine

Cite this

Optimization of alarms : A study on alarm limits, alarm sounds, and false alarms, intended to reduce annoyance. / Block, Frank E; Nuutinen, Lauri; Ballast, Bert.

In: Journal of Clinical Monitoring and Computing, Vol. 15, No. 2, 26.06.1999, p. 75-83.

Research output: Contribution to journalArticle

@article{9634679a01b24773b4f993ed41c7f5cd,
title = "Optimization of alarms: A study on alarm limits, alarm sounds, and false alarms, intended to reduce annoyance",
abstract = "Alarms in the operating room remain a major source of annoyance and confusion. Nearly all alarms result from a transgression of certain alarm limits. We surveyed manufacturers at a major meeting of anesthesiologists for their default alarm limits. We also conducted a mail survey of anesthesiologists in the United States, Finland, and the Netherlands, to learn their chosen alarm limits, reasons for turning off the alarms, and estimates of the number of false alarms. The surveys of medical monitoring manufacturers demonstrated a wide variety of default alarm settings. Anesthesiologists' chosen alarm limits tended to parallel the manufacturers' defaults. Fewer than 30{\%} of anesthesiologists stated that they did not turn off the alarms. The leading reason for turning off alarms was the large number of false alarms. Estimates of the number of false alarms varied from 30{\%} to 76{\%}. The desired role of alarms in anesthetic practice and some suggestions for the general improvement of alarms are discussed.",
author = "Block, {Frank E} and Lauri Nuutinen and Bert Ballast",
year = "1999",
month = "6",
day = "26",
doi = "10.1023/A:1009992830942",
language = "English (US)",
volume = "15",
pages = "75--83",
journal = "Journal of Clinical Monitoring and Computing",
issn = "1387-1307",
publisher = "Springer Netherlands",
number = "2",

}

TY - JOUR

T1 - Optimization of alarms

T2 - A study on alarm limits, alarm sounds, and false alarms, intended to reduce annoyance

AU - Block, Frank E

AU - Nuutinen, Lauri

AU - Ballast, Bert

PY - 1999/6/26

Y1 - 1999/6/26

N2 - Alarms in the operating room remain a major source of annoyance and confusion. Nearly all alarms result from a transgression of certain alarm limits. We surveyed manufacturers at a major meeting of anesthesiologists for their default alarm limits. We also conducted a mail survey of anesthesiologists in the United States, Finland, and the Netherlands, to learn their chosen alarm limits, reasons for turning off the alarms, and estimates of the number of false alarms. The surveys of medical monitoring manufacturers demonstrated a wide variety of default alarm settings. Anesthesiologists' chosen alarm limits tended to parallel the manufacturers' defaults. Fewer than 30% of anesthesiologists stated that they did not turn off the alarms. The leading reason for turning off alarms was the large number of false alarms. Estimates of the number of false alarms varied from 30% to 76%. The desired role of alarms in anesthetic practice and some suggestions for the general improvement of alarms are discussed.

AB - Alarms in the operating room remain a major source of annoyance and confusion. Nearly all alarms result from a transgression of certain alarm limits. We surveyed manufacturers at a major meeting of anesthesiologists for their default alarm limits. We also conducted a mail survey of anesthesiologists in the United States, Finland, and the Netherlands, to learn their chosen alarm limits, reasons for turning off the alarms, and estimates of the number of false alarms. The surveys of medical monitoring manufacturers demonstrated a wide variety of default alarm settings. Anesthesiologists' chosen alarm limits tended to parallel the manufacturers' defaults. Fewer than 30% of anesthesiologists stated that they did not turn off the alarms. The leading reason for turning off alarms was the large number of false alarms. Estimates of the number of false alarms varied from 30% to 76%. The desired role of alarms in anesthetic practice and some suggestions for the general improvement of alarms are discussed.

UR - http://www.scopus.com/inward/record.url?scp=0344654720&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0344654720&partnerID=8YFLogxK

U2 - 10.1023/A:1009992830942

DO - 10.1023/A:1009992830942

M3 - Article

VL - 15

SP - 75

EP - 83

JO - Journal of Clinical Monitoring and Computing

JF - Journal of Clinical Monitoring and Computing

SN - 1387-1307

IS - 2

ER -