Periodic Variations in Transcranial Doppler Mean Velocities

Cole A. Giller, Michael Lam, Angela Roseland

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

Although serial transcranial Doppler measurements of blood flow velocity are of considerable clinical utility, their use assumes that the velocity signals are stable and unchanging during short time periods. Contrary to this assumption, the authors found significant variations in mean velocity signals in both normal subjects and patients, which may confound the interpretation of serial studies. Signals were continuously obtained for 5 to 1 0 minutes from the middle cerebral artery of 11 normal subjects and 18 patients (22 studies) with a variety of neurosurgical disorders. The average difference between the peak and the trough in the waves observed in the normal population was 11 ± 4% (standard error of mean), and 5 of the 11 had at least one wave with a difference of more than 20%. The average difference between the peak and trough signals in the neurosurgical population was 14 ± 13% (SEM) and 13 of the 22 studies showed at least one wave with a difference of more than 20%. These variations were consistently seen and may be related to similar variations in blood pressure or intracranial pressure waves. Whatever the origin, these variations should be recognized during the interpretation of transcranial Doppler signals in clinical practice.

Original languageEnglish (US)
Pages (from-to)160-162
Number of pages3
JournalJournal of Neuroimaging
Volume3
Issue number3
DOIs
StatePublished - Jul 1 1993
Externally publishedYes

Fingerprint

Blood Flow Velocity
Middle Cerebral Artery
Intracranial Pressure
Population
Blood Pressure

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Clinical Neurology
  • Neuroscience(all)
  • Radiological and Ultrasound Technology

Cite this

Periodic Variations in Transcranial Doppler Mean Velocities. / Giller, Cole A.; Lam, Michael; Roseland, Angela.

In: Journal of Neuroimaging, Vol. 3, No. 3, 01.07.1993, p. 160-162.

Research output: Contribution to journalArticle

Giller, Cole A. ; Lam, Michael ; Roseland, Angela. / Periodic Variations in Transcranial Doppler Mean Velocities. In: Journal of Neuroimaging. 1993 ; Vol. 3, No. 3. pp. 160-162.
@article{288a6c7975b644b98671c73f510258ff,
title = "Periodic Variations in Transcranial Doppler Mean Velocities",
abstract = "Although serial transcranial Doppler measurements of blood flow velocity are of considerable clinical utility, their use assumes that the velocity signals are stable and unchanging during short time periods. Contrary to this assumption, the authors found significant variations in mean velocity signals in both normal subjects and patients, which may confound the interpretation of serial studies. Signals were continuously obtained for 5 to 1 0 minutes from the middle cerebral artery of 11 normal subjects and 18 patients (22 studies) with a variety of neurosurgical disorders. The average difference between the peak and the trough in the waves observed in the normal population was 11 ± 4{\%} (standard error of mean), and 5 of the 11 had at least one wave with a difference of more than 20{\%}. The average difference between the peak and trough signals in the neurosurgical population was 14 ± 13{\%} (SEM) and 13 of the 22 studies showed at least one wave with a difference of more than 20{\%}. These variations were consistently seen and may be related to similar variations in blood pressure or intracranial pressure waves. Whatever the origin, these variations should be recognized during the interpretation of transcranial Doppler signals in clinical practice.",
author = "Giller, {Cole A.} and Michael Lam and Angela Roseland",
year = "1993",
month = "7",
day = "1",
doi = "10.1111/jon199333160",
language = "English (US)",
volume = "3",
pages = "160--162",
journal = "Journal of Neuroimaging",
issn = "1051-2284",
publisher = "Wiley-Blackwell",
number = "3",

}

TY - JOUR

T1 - Periodic Variations in Transcranial Doppler Mean Velocities

AU - Giller, Cole A.

AU - Lam, Michael

AU - Roseland, Angela

PY - 1993/7/1

Y1 - 1993/7/1

N2 - Although serial transcranial Doppler measurements of blood flow velocity are of considerable clinical utility, their use assumes that the velocity signals are stable and unchanging during short time periods. Contrary to this assumption, the authors found significant variations in mean velocity signals in both normal subjects and patients, which may confound the interpretation of serial studies. Signals were continuously obtained for 5 to 1 0 minutes from the middle cerebral artery of 11 normal subjects and 18 patients (22 studies) with a variety of neurosurgical disorders. The average difference between the peak and the trough in the waves observed in the normal population was 11 ± 4% (standard error of mean), and 5 of the 11 had at least one wave with a difference of more than 20%. The average difference between the peak and trough signals in the neurosurgical population was 14 ± 13% (SEM) and 13 of the 22 studies showed at least one wave with a difference of more than 20%. These variations were consistently seen and may be related to similar variations in blood pressure or intracranial pressure waves. Whatever the origin, these variations should be recognized during the interpretation of transcranial Doppler signals in clinical practice.

AB - Although serial transcranial Doppler measurements of blood flow velocity are of considerable clinical utility, their use assumes that the velocity signals are stable and unchanging during short time periods. Contrary to this assumption, the authors found significant variations in mean velocity signals in both normal subjects and patients, which may confound the interpretation of serial studies. Signals were continuously obtained for 5 to 1 0 minutes from the middle cerebral artery of 11 normal subjects and 18 patients (22 studies) with a variety of neurosurgical disorders. The average difference between the peak and the trough in the waves observed in the normal population was 11 ± 4% (standard error of mean), and 5 of the 11 had at least one wave with a difference of more than 20%. The average difference between the peak and trough signals in the neurosurgical population was 14 ± 13% (SEM) and 13 of the 22 studies showed at least one wave with a difference of more than 20%. These variations were consistently seen and may be related to similar variations in blood pressure or intracranial pressure waves. Whatever the origin, these variations should be recognized during the interpretation of transcranial Doppler signals in clinical practice.

UR - http://www.scopus.com/inward/record.url?scp=0027209197&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027209197&partnerID=8YFLogxK

U2 - 10.1111/jon199333160

DO - 10.1111/jon199333160

M3 - Article

C2 - 10150150

AN - SCOPUS:0027209197

VL - 3

SP - 160

EP - 162

JO - Journal of Neuroimaging

JF - Journal of Neuroimaging

SN - 1051-2284

IS - 3

ER -