Preliminary investigation and application of a novel deformable PRESAGE® dosimeter

T. Juang, Joseph R Newton, S. Das, J. Adamovics, M. Oldham

Research output: Contribution to journalConference article

5 Citations (Scopus)

Abstract

Deformable 3D dosimeters have potential applications in validating deformable dose mapping algorithms. This study evaluates a novel deformable PRESAGE® dosimeter and its application toward validating the deformable algorithm employed by VelocityAI. The deformable PRESAGE® dosimeter exhibited a linear dose response with a sensitivity of 0.0032 ΔOD/(Gy/cm). Comparison of an experimental dosimeter irradiated with an MLC pencilbeam checkerboard pattern under lateral compression up to 27% to a non-deformed control dosimeter irradiated with the same pattern verified dose tracking under deformation. CTs of the experimental dosimeter prior to and during compression were exported into VelocityAI and used to map an Eclipse dose distribution calculated on the compressed dosimeter to its original shape. A comparison between the VelocityAI dose distribution and the distribution from the dosimeter showed field displacements up to 7.3 mm and up to a 175% difference in field dimensions. These results highlight the need for validating deformable dose mapping algorithms to ensure patient safety and quality of care.

Original languageEnglish (US)
Article number012080
JournalJournal of Physics: Conference Series
Volume444
Issue number1
DOIs
StatePublished - Jan 1 2013
Externally publishedYes
Event7th International Conference on 3D Radiation Dosimetry, IC3DDose 2012 - Sydney, NSW, Australia
Duration: Dec 4 2012Dec 8 2012

Fingerprint

dosimeters
dosage
eclipses
safety
sensitivity

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Cite this

Preliminary investigation and application of a novel deformable PRESAGE® dosimeter. / Juang, T.; Newton, Joseph R; Das, S.; Adamovics, J.; Oldham, M.

In: Journal of Physics: Conference Series, Vol. 444, No. 1, 012080, 01.01.2013.

Research output: Contribution to journalConference article

@article{508a0cd8c9ad4524a68d642bf8a96d41,
title = "Preliminary investigation and application of a novel deformable PRESAGE{\circledR} dosimeter",
abstract = "Deformable 3D dosimeters have potential applications in validating deformable dose mapping algorithms. This study evaluates a novel deformable PRESAGE{\circledR} dosimeter and its application toward validating the deformable algorithm employed by VelocityAI. The deformable PRESAGE{\circledR} dosimeter exhibited a linear dose response with a sensitivity of 0.0032 ΔOD/(Gy/cm). Comparison of an experimental dosimeter irradiated with an MLC pencilbeam checkerboard pattern under lateral compression up to 27{\%} to a non-deformed control dosimeter irradiated with the same pattern verified dose tracking under deformation. CTs of the experimental dosimeter prior to and during compression were exported into VelocityAI and used to map an Eclipse dose distribution calculated on the compressed dosimeter to its original shape. A comparison between the VelocityAI dose distribution and the distribution from the dosimeter showed field displacements up to 7.3 mm and up to a 175{\%} difference in field dimensions. These results highlight the need for validating deformable dose mapping algorithms to ensure patient safety and quality of care.",
author = "T. Juang and Newton, {Joseph R} and S. Das and J. Adamovics and M. Oldham",
year = "2013",
month = "1",
day = "1",
doi = "10.1088/1742-6596/444/1/012080",
language = "English (US)",
volume = "444",
journal = "Journal of Physics: Conference Series",
issn = "1742-6588",
publisher = "IOP Publishing Ltd.",
number = "1",

}

TY - JOUR

T1 - Preliminary investigation and application of a novel deformable PRESAGE® dosimeter

AU - Juang, T.

AU - Newton, Joseph R

AU - Das, S.

AU - Adamovics, J.

AU - Oldham, M.

PY - 2013/1/1

Y1 - 2013/1/1

N2 - Deformable 3D dosimeters have potential applications in validating deformable dose mapping algorithms. This study evaluates a novel deformable PRESAGE® dosimeter and its application toward validating the deformable algorithm employed by VelocityAI. The deformable PRESAGE® dosimeter exhibited a linear dose response with a sensitivity of 0.0032 ΔOD/(Gy/cm). Comparison of an experimental dosimeter irradiated with an MLC pencilbeam checkerboard pattern under lateral compression up to 27% to a non-deformed control dosimeter irradiated with the same pattern verified dose tracking under deformation. CTs of the experimental dosimeter prior to and during compression were exported into VelocityAI and used to map an Eclipse dose distribution calculated on the compressed dosimeter to its original shape. A comparison between the VelocityAI dose distribution and the distribution from the dosimeter showed field displacements up to 7.3 mm and up to a 175% difference in field dimensions. These results highlight the need for validating deformable dose mapping algorithms to ensure patient safety and quality of care.

AB - Deformable 3D dosimeters have potential applications in validating deformable dose mapping algorithms. This study evaluates a novel deformable PRESAGE® dosimeter and its application toward validating the deformable algorithm employed by VelocityAI. The deformable PRESAGE® dosimeter exhibited a linear dose response with a sensitivity of 0.0032 ΔOD/(Gy/cm). Comparison of an experimental dosimeter irradiated with an MLC pencilbeam checkerboard pattern under lateral compression up to 27% to a non-deformed control dosimeter irradiated with the same pattern verified dose tracking under deformation. CTs of the experimental dosimeter prior to and during compression were exported into VelocityAI and used to map an Eclipse dose distribution calculated on the compressed dosimeter to its original shape. A comparison between the VelocityAI dose distribution and the distribution from the dosimeter showed field displacements up to 7.3 mm and up to a 175% difference in field dimensions. These results highlight the need for validating deformable dose mapping algorithms to ensure patient safety and quality of care.

UR - http://www.scopus.com/inward/record.url?scp=84883391232&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84883391232&partnerID=8YFLogxK

U2 - 10.1088/1742-6596/444/1/012080

DO - 10.1088/1742-6596/444/1/012080

M3 - Conference article

VL - 444

JO - Journal of Physics: Conference Series

JF - Journal of Physics: Conference Series

SN - 1742-6588

IS - 1

M1 - 012080

ER -