Pulse oximetry: Principles and limitations

Research output: Contribution to journalReview article

248 Citations (Scopus)

Abstract

The pulse oximeter has become an essential tool in the modern practice of emergency medicine. However, despite the reliance placed on the information this monitor offers, the underlying principles and associated limitations of pulse oximetry are poorly understood by medical practitioners. This article reviews the principles of pulse oximetry, with an eye toward recognizing the limitations of this tool. Among these are performance limitations in the settings of carboxyhemoglobinemia, methemoglobinemia, motion artifact, hypotension, vasoconstriction, and anemia. The accuracy of pulse oximetry is discussed in light of these factors, with further discussion of applications for pulse oximetry in emergency medicine, including both oximetric and plethysmographic operation. The pulse oximeter is an invaluable instrument for emergency medicine practice, but as with any test the data it offers must be critically appraised for proper interpretation and utilization.

Original languageEnglish (US)
Pages (from-to)59-66
Number of pages8
JournalAmerican Journal of Emergency Medicine
Volume17
Issue number1
DOIs
StatePublished - Jan 1 1999

Fingerprint

Oximetry
Emergency Medicine
Pulse
Methemoglobinemia
Vasoconstriction
Hypotension
Artifacts
Anemia

Keywords

  • Anoxia
  • Oximetry
  • Physiologic monitoring
  • Plethysmography

ASJC Scopus subject areas

  • Emergency Medicine

Cite this

Pulse oximetry : Principles and limitations. / Sinex, James E.

In: American Journal of Emergency Medicine, Vol. 17, No. 1, 01.01.1999, p. 59-66.

Research output: Contribution to journalReview article

@article{1a2b926984f5453d881c9e17de9082f0,
title = "Pulse oximetry: Principles and limitations",
abstract = "The pulse oximeter has become an essential tool in the modern practice of emergency medicine. However, despite the reliance placed on the information this monitor offers, the underlying principles and associated limitations of pulse oximetry are poorly understood by medical practitioners. This article reviews the principles of pulse oximetry, with an eye toward recognizing the limitations of this tool. Among these are performance limitations in the settings of carboxyhemoglobinemia, methemoglobinemia, motion artifact, hypotension, vasoconstriction, and anemia. The accuracy of pulse oximetry is discussed in light of these factors, with further discussion of applications for pulse oximetry in emergency medicine, including both oximetric and plethysmographic operation. The pulse oximeter is an invaluable instrument for emergency medicine practice, but as with any test the data it offers must be critically appraised for proper interpretation and utilization.",
keywords = "Anoxia, Oximetry, Physiologic monitoring, Plethysmography",
author = "Sinex, {James E}",
year = "1999",
month = "1",
day = "1",
doi = "10.1016/S0735-6757(99)90019-0",
language = "English (US)",
volume = "17",
pages = "59--66",
journal = "American Journal of Emergency Medicine",
issn = "0735-6757",
publisher = "W.B. Saunders Ltd",
number = "1",

}

TY - JOUR

T1 - Pulse oximetry

T2 - Principles and limitations

AU - Sinex, James E

PY - 1999/1/1

Y1 - 1999/1/1

N2 - The pulse oximeter has become an essential tool in the modern practice of emergency medicine. However, despite the reliance placed on the information this monitor offers, the underlying principles and associated limitations of pulse oximetry are poorly understood by medical practitioners. This article reviews the principles of pulse oximetry, with an eye toward recognizing the limitations of this tool. Among these are performance limitations in the settings of carboxyhemoglobinemia, methemoglobinemia, motion artifact, hypotension, vasoconstriction, and anemia. The accuracy of pulse oximetry is discussed in light of these factors, with further discussion of applications for pulse oximetry in emergency medicine, including both oximetric and plethysmographic operation. The pulse oximeter is an invaluable instrument for emergency medicine practice, but as with any test the data it offers must be critically appraised for proper interpretation and utilization.

AB - The pulse oximeter has become an essential tool in the modern practice of emergency medicine. However, despite the reliance placed on the information this monitor offers, the underlying principles and associated limitations of pulse oximetry are poorly understood by medical practitioners. This article reviews the principles of pulse oximetry, with an eye toward recognizing the limitations of this tool. Among these are performance limitations in the settings of carboxyhemoglobinemia, methemoglobinemia, motion artifact, hypotension, vasoconstriction, and anemia. The accuracy of pulse oximetry is discussed in light of these factors, with further discussion of applications for pulse oximetry in emergency medicine, including both oximetric and plethysmographic operation. The pulse oximeter is an invaluable instrument for emergency medicine practice, but as with any test the data it offers must be critically appraised for proper interpretation and utilization.

KW - Anoxia

KW - Oximetry

KW - Physiologic monitoring

KW - Plethysmography

UR - http://www.scopus.com/inward/record.url?scp=0032932450&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0032932450&partnerID=8YFLogxK

U2 - 10.1016/S0735-6757(99)90019-0

DO - 10.1016/S0735-6757(99)90019-0

M3 - Review article

C2 - 9928703

AN - SCOPUS:0032932450

VL - 17

SP - 59

EP - 66

JO - American Journal of Emergency Medicine

JF - American Journal of Emergency Medicine

SN - 0735-6757

IS - 1

ER -