SU‐FF‐T‐391: Study of Treatment Couch Leveling and Sagging Related to Patient Weight and Couch Longitudinal Positions

Q. Chen, JianYue Jin, H. Guan, R. Hammoud, S. Sheim, S. li, S. Ryu, B. Movsas

Research output: Contribution to journalArticle

Abstract

Purpose: The weighted couch appears leveling and sagging inheriting from its design and material used. The amount of leveling and sagging depend on longitudinal position and patient weight since the couch experiences different amounts of torque. This phenomenon could affect patient setup accuracy, especially for stereotactic radiosurgery. Material and Method: An infrared camera based ExacTrac system (BrainLab) of high accuracy was used to measure the leveling and sagging. Four infrared reflecting makers were placed on a piece of graphic paper attached on the couch‐top, with a 10 cm separation in superior/inferior direction, and 6cm in lateral direction. Couch base is ExactCouch (Varian) with a new fiberglass couch top (Brainlab). The measurements were taken at a 5cm increment step on longitudinal direction in a range of 80cm couch movement, repeated with weight loads of 40LB and 60 LB on the couch front end. The graphic paper was repositioned for each couch movement so that the center of the four markers was always at the isocenter. The average horizontal position change of four markers represented the sagging, and the position difference between the superior and inferior markers represented the leveling. Result: Couch sagging was Z(mm) = 0.011*Y(cm) + 0.26 for zero load, Z(mm) = −0.0048*Y(cm) + 0.08 for 40Lb, and Z(mm) = −0.011Y(cm) + 0.03 for 60Lb, where Y is the amount of couch longitudinal movement. The maximum sagging was 1.9mm compared the difference between 60Lb loading and 0 loading. Couch leveling was 0.316±0.084mm, 0.58±0.12mm, 0.71±0.059mm, for three different loading with the range of 11 cm around the isocenter, however, it changed little with different couch position. Conclusion: Couch sagging depends on its longitudinal extension and the patient weight. The maximum sagging at isocenter position was about 2mm. Couch leveling only depend on the weight of loading.

Original languageEnglish (US)
Pages (from-to)2491-2492
Number of pages2
JournalMedical Physics
Volume34
Issue number6
DOIs
StatePublished - Jan 1 2007

Fingerprint

Weights and Measures
Radiosurgery
Torque
Therapeutics
Direction compound
fiberglass

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

SU‐FF‐T‐391 : Study of Treatment Couch Leveling and Sagging Related to Patient Weight and Couch Longitudinal Positions. / Chen, Q.; Jin, JianYue; Guan, H.; Hammoud, R.; Sheim, S.; li, S.; Ryu, S.; Movsas, B.

In: Medical Physics, Vol. 34, No. 6, 01.01.2007, p. 2491-2492.

Research output: Contribution to journalArticle

Chen, Q, Jin, J, Guan, H, Hammoud, R, Sheim, S, li, S, Ryu, S & Movsas, B 2007, 'SU‐FF‐T‐391: Study of Treatment Couch Leveling and Sagging Related to Patient Weight and Couch Longitudinal Positions', Medical Physics, vol. 34, no. 6, pp. 2491-2492. https://doi.org/10.1118/1.2761116
Chen, Q. ; Jin, JianYue ; Guan, H. ; Hammoud, R. ; Sheim, S. ; li, S. ; Ryu, S. ; Movsas, B. / SU‐FF‐T‐391 : Study of Treatment Couch Leveling and Sagging Related to Patient Weight and Couch Longitudinal Positions. In: Medical Physics. 2007 ; Vol. 34, No. 6. pp. 2491-2492.
@article{b2214d0718734661b818ad2640700ef3,
title = "SU‐FF‐T‐391: Study of Treatment Couch Leveling and Sagging Related to Patient Weight and Couch Longitudinal Positions",
abstract = "Purpose: The weighted couch appears leveling and sagging inheriting from its design and material used. The amount of leveling and sagging depend on longitudinal position and patient weight since the couch experiences different amounts of torque. This phenomenon could affect patient setup accuracy, especially for stereotactic radiosurgery. Material and Method: An infrared camera based ExacTrac system (BrainLab) of high accuracy was used to measure the leveling and sagging. Four infrared reflecting makers were placed on a piece of graphic paper attached on the couch‐top, with a 10 cm separation in superior/inferior direction, and 6cm in lateral direction. Couch base is ExactCouch (Varian) with a new fiberglass couch top (Brainlab). The measurements were taken at a 5cm increment step on longitudinal direction in a range of 80cm couch movement, repeated with weight loads of 40LB and 60 LB on the couch front end. The graphic paper was repositioned for each couch movement so that the center of the four markers was always at the isocenter. The average horizontal position change of four markers represented the sagging, and the position difference between the superior and inferior markers represented the leveling. Result: Couch sagging was Z(mm) = 0.011*Y(cm) + 0.26 for zero load, Z(mm) = −0.0048*Y(cm) + 0.08 for 40Lb, and Z(mm) = −0.011Y(cm) + 0.03 for 60Lb, where Y is the amount of couch longitudinal movement. The maximum sagging was 1.9mm compared the difference between 60Lb loading and 0 loading. Couch leveling was 0.316±0.084mm, 0.58±0.12mm, 0.71±0.059mm, for three different loading with the range of 11 cm around the isocenter, however, it changed little with different couch position. Conclusion: Couch sagging depends on its longitudinal extension and the patient weight. The maximum sagging at isocenter position was about 2mm. Couch leveling only depend on the weight of loading.",
author = "Q. Chen and JianYue Jin and H. Guan and R. Hammoud and S. Sheim and S. li and S. Ryu and B. Movsas",
year = "2007",
month = "1",
day = "1",
doi = "10.1118/1.2761116",
language = "English (US)",
volume = "34",
pages = "2491--2492",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - SU‐FF‐T‐391

T2 - Study of Treatment Couch Leveling and Sagging Related to Patient Weight and Couch Longitudinal Positions

AU - Chen, Q.

AU - Jin, JianYue

AU - Guan, H.

AU - Hammoud, R.

AU - Sheim, S.

AU - li, S.

AU - Ryu, S.

AU - Movsas, B.

PY - 2007/1/1

Y1 - 2007/1/1

N2 - Purpose: The weighted couch appears leveling and sagging inheriting from its design and material used. The amount of leveling and sagging depend on longitudinal position and patient weight since the couch experiences different amounts of torque. This phenomenon could affect patient setup accuracy, especially for stereotactic radiosurgery. Material and Method: An infrared camera based ExacTrac system (BrainLab) of high accuracy was used to measure the leveling and sagging. Four infrared reflecting makers were placed on a piece of graphic paper attached on the couch‐top, with a 10 cm separation in superior/inferior direction, and 6cm in lateral direction. Couch base is ExactCouch (Varian) with a new fiberglass couch top (Brainlab). The measurements were taken at a 5cm increment step on longitudinal direction in a range of 80cm couch movement, repeated with weight loads of 40LB and 60 LB on the couch front end. The graphic paper was repositioned for each couch movement so that the center of the four markers was always at the isocenter. The average horizontal position change of four markers represented the sagging, and the position difference between the superior and inferior markers represented the leveling. Result: Couch sagging was Z(mm) = 0.011*Y(cm) + 0.26 for zero load, Z(mm) = −0.0048*Y(cm) + 0.08 for 40Lb, and Z(mm) = −0.011Y(cm) + 0.03 for 60Lb, where Y is the amount of couch longitudinal movement. The maximum sagging was 1.9mm compared the difference between 60Lb loading and 0 loading. Couch leveling was 0.316±0.084mm, 0.58±0.12mm, 0.71±0.059mm, for three different loading with the range of 11 cm around the isocenter, however, it changed little with different couch position. Conclusion: Couch sagging depends on its longitudinal extension and the patient weight. The maximum sagging at isocenter position was about 2mm. Couch leveling only depend on the weight of loading.

AB - Purpose: The weighted couch appears leveling and sagging inheriting from its design and material used. The amount of leveling and sagging depend on longitudinal position and patient weight since the couch experiences different amounts of torque. This phenomenon could affect patient setup accuracy, especially for stereotactic radiosurgery. Material and Method: An infrared camera based ExacTrac system (BrainLab) of high accuracy was used to measure the leveling and sagging. Four infrared reflecting makers were placed on a piece of graphic paper attached on the couch‐top, with a 10 cm separation in superior/inferior direction, and 6cm in lateral direction. Couch base is ExactCouch (Varian) with a new fiberglass couch top (Brainlab). The measurements were taken at a 5cm increment step on longitudinal direction in a range of 80cm couch movement, repeated with weight loads of 40LB and 60 LB on the couch front end. The graphic paper was repositioned for each couch movement so that the center of the four markers was always at the isocenter. The average horizontal position change of four markers represented the sagging, and the position difference between the superior and inferior markers represented the leveling. Result: Couch sagging was Z(mm) = 0.011*Y(cm) + 0.26 for zero load, Z(mm) = −0.0048*Y(cm) + 0.08 for 40Lb, and Z(mm) = −0.011Y(cm) + 0.03 for 60Lb, where Y is the amount of couch longitudinal movement. The maximum sagging was 1.9mm compared the difference between 60Lb loading and 0 loading. Couch leveling was 0.316±0.084mm, 0.58±0.12mm, 0.71±0.059mm, for three different loading with the range of 11 cm around the isocenter, however, it changed little with different couch position. Conclusion: Couch sagging depends on its longitudinal extension and the patient weight. The maximum sagging at isocenter position was about 2mm. Couch leveling only depend on the weight of loading.

UR - http://www.scopus.com/inward/record.url?scp=85024775223&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024775223&partnerID=8YFLogxK

U2 - 10.1118/1.2761116

DO - 10.1118/1.2761116

M3 - Article

AN - SCOPUS:85024775223

VL - 34

SP - 2491

EP - 2492

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -