We studied the function of four different monitoring probes used with the Satlite pulse oximeter (Datex, Helsinki). The aim was to evaluate ease of use and compare the function of the probes and their attachment methods in different locations (finger, toe, ear, thumb web, instep, wrist, and ankle). Two similar pulse wave oximeters were used in the study. To select the best signal we determined the absolute height of the plethysmographic waveform for each probe. Probes were compared on awake normal adult volunteers (N=13), in anesthetized adult patients (N=12), and in neonates weighting 500-1000 g (N=8). In all the adult trials, the clip-on finger probe was used as a reference, and probes taped or clipped onto the finger provided adequate signals in comparison. Taped-on probes gave also satisfactory results on the toe. It was difficult to get a quality signal from the ear or from the thumb web, however. In the neonates, taped-on probes were most satisfactory.

Original languageEnglish (US)
Pages (from-to)23-28
Number of pages6
JournalInternational Journal of Clinical Monitoring and Computing
Volume14
Issue number1
StatePublished - Dec 1 1997
Externally publishedYes

Fingerprint

Fingers
Thumb
Toes
Technology
Ear
Newborn Infant
Wrist
Surgical Instruments
Ankle
Healthy Volunteers

Keywords

  • Adults
  • Cutaneous
  • Monitoring
  • Neonates
  • Oximetries
  • Probe
  • Pulse oximetry
  • Signal strength

ASJC Scopus subject areas

  • Medicine (miscellaneous)
  • Critical Care and Intensive Care Medicine

Cite this

Technology évaluation report : Obtaining pulse oximeter signals when the usual probe cannot be used. / Block, Frank E; Fuhrman, Thomas M.; Cordero, Leandro; Schaaf, Carl.

In: International Journal of Clinical Monitoring and Computing, Vol. 14, No. 1, 01.12.1997, p. 23-28.

Research output: Contribution to journalArticle

Block, Frank E ; Fuhrman, Thomas M. ; Cordero, Leandro ; Schaaf, Carl. / Technology évaluation report : Obtaining pulse oximeter signals when the usual probe cannot be used. In: International Journal of Clinical Monitoring and Computing. 1997 ; Vol. 14, No. 1. pp. 23-28.
@article{14847dc297e54f899778e11247ec3e75,
title = "Technology {\'e}valuation report: Obtaining pulse oximeter signals when the usual probe cannot be used",
abstract = "We studied the function of four different monitoring probes used with the Satlite pulse oximeter (Datex, Helsinki). The aim was to evaluate ease of use and compare the function of the probes and their attachment methods in different locations (finger, toe, ear, thumb web, instep, wrist, and ankle). Two similar pulse wave oximeters were used in the study. To select the best signal we determined the absolute height of the plethysmographic waveform for each probe. Probes were compared on awake normal adult volunteers (N=13), in anesthetized adult patients (N=12), and in neonates weighting 500-1000 g (N=8). In all the adult trials, the clip-on finger probe was used as a reference, and probes taped or clipped onto the finger provided adequate signals in comparison. Taped-on probes gave also satisfactory results on the toe. It was difficult to get a quality signal from the ear or from the thumb web, however. In the neonates, taped-on probes were most satisfactory.",
keywords = "Adults, Cutaneous, Monitoring, Neonates, Oximetries, Probe, Pulse oximetry, Signal strength",
author = "Block, {Frank E} and Fuhrman, {Thomas M.} and Leandro Cordero and Carl Schaaf",
year = "1997",
month = "12",
day = "1",
language = "English (US)",
volume = "14",
pages = "23--28",
journal = "Journal of Clinical Monitoring and Computing",
issn = "1387-1307",
publisher = "Springer Netherlands",
number = "1",

}

TY - JOUR

T1 - Technology évaluation report

T2 - Obtaining pulse oximeter signals when the usual probe cannot be used

AU - Block, Frank E

AU - Fuhrman, Thomas M.

AU - Cordero, Leandro

AU - Schaaf, Carl

PY - 1997/12/1

Y1 - 1997/12/1

N2 - We studied the function of four different monitoring probes used with the Satlite pulse oximeter (Datex, Helsinki). The aim was to evaluate ease of use and compare the function of the probes and their attachment methods in different locations (finger, toe, ear, thumb web, instep, wrist, and ankle). Two similar pulse wave oximeters were used in the study. To select the best signal we determined the absolute height of the plethysmographic waveform for each probe. Probes were compared on awake normal adult volunteers (N=13), in anesthetized adult patients (N=12), and in neonates weighting 500-1000 g (N=8). In all the adult trials, the clip-on finger probe was used as a reference, and probes taped or clipped onto the finger provided adequate signals in comparison. Taped-on probes gave also satisfactory results on the toe. It was difficult to get a quality signal from the ear or from the thumb web, however. In the neonates, taped-on probes were most satisfactory.

AB - We studied the function of four different monitoring probes used with the Satlite pulse oximeter (Datex, Helsinki). The aim was to evaluate ease of use and compare the function of the probes and their attachment methods in different locations (finger, toe, ear, thumb web, instep, wrist, and ankle). Two similar pulse wave oximeters were used in the study. To select the best signal we determined the absolute height of the plethysmographic waveform for each probe. Probes were compared on awake normal adult volunteers (N=13), in anesthetized adult patients (N=12), and in neonates weighting 500-1000 g (N=8). In all the adult trials, the clip-on finger probe was used as a reference, and probes taped or clipped onto the finger provided adequate signals in comparison. Taped-on probes gave also satisfactory results on the toe. It was difficult to get a quality signal from the ear or from the thumb web, however. In the neonates, taped-on probes were most satisfactory.

KW - Adults

KW - Cutaneous

KW - Monitoring

KW - Neonates

KW - Oximetries

KW - Probe

KW - Pulse oximetry

KW - Signal strength

UR - http://www.scopus.com/inward/record.url?scp=0030902153&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030902153&partnerID=8YFLogxK

M3 - Article

VL - 14

SP - 23

EP - 28

JO - Journal of Clinical Monitoring and Computing

JF - Journal of Clinical Monitoring and Computing

SN - 1387-1307

IS - 1

ER -