The AM5600 ambulatory blood pressure recording system

Phyllis A. Richey, Curtis L. Jones, Gregory A Harshfield, Grant W. Somes, Karen C. Johnson, James E. Bailey, Judith E. Soberman

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

Objective. To describe our experience with the AM5600 (Advanced Medical Products, Inc.) ambulatory blood pressure recorder-Holter monitor which has the unique features of providing full disclosure Hotter data and the storage of Korotkoff signals for each blood pressure measurement. Method. We performed a total of 236 recording sessions with 118 normotensive subjects aged 49-74 years. We then selected 25 of the recordings randomly and examined them on two different occasions, first with the aid of the Korotkoff signals and later using the traditional technique, without the aid of the Korotkoff signals. Result. Two hundred and sixteen (91.5%) recordings were completed successfully. Within these data sets, on average 82.3 ± 12.1% of the measurements per recording were considered acceptable after examination of the Korotkoff signals. This comparison between the traditional and the Korotkoff technique resulted in a sensitivity 11.3% and a specificity 99.2% of those of the traditional method. Conclusion. The performance of the AM5600 is satisfactory and the storage of Korotkoff signals improves the detection of artifacts.

Original languageEnglish (US)
Pages (from-to)193-195
Number of pages3
JournalBlood Pressure Monitoring
Volume2
Issue number3-4
StatePublished - Jan 1 1997
Externally publishedYes

Fingerprint

Blood Pressure
First Aid
Information Storage and Retrieval
Disclosure
Artifacts
Datasets
Psychological Signal Detection

Keywords

  • Ambulatory blood pressure monitoring
  • Auscultatory
  • Korotkoff signal

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine
  • Internal Medicine

Cite this

Richey, P. A., Jones, C. L., Harshfield, G. A., Somes, G. W., Johnson, K. C., Bailey, J. E., & Soberman, J. E. (1997). The AM5600 ambulatory blood pressure recording system. Blood Pressure Monitoring, 2(3-4), 193-195.

The AM5600 ambulatory blood pressure recording system. / Richey, Phyllis A.; Jones, Curtis L.; Harshfield, Gregory A; Somes, Grant W.; Johnson, Karen C.; Bailey, James E.; Soberman, Judith E.

In: Blood Pressure Monitoring, Vol. 2, No. 3-4, 01.01.1997, p. 193-195.

Research output: Contribution to journalArticle

Richey, PA, Jones, CL, Harshfield, GA, Somes, GW, Johnson, KC, Bailey, JE & Soberman, JE 1997, 'The AM5600 ambulatory blood pressure recording system', Blood Pressure Monitoring, vol. 2, no. 3-4, pp. 193-195.
Richey PA, Jones CL, Harshfield GA, Somes GW, Johnson KC, Bailey JE et al. The AM5600 ambulatory blood pressure recording system. Blood Pressure Monitoring. 1997 Jan 1;2(3-4):193-195.
Richey, Phyllis A. ; Jones, Curtis L. ; Harshfield, Gregory A ; Somes, Grant W. ; Johnson, Karen C. ; Bailey, James E. ; Soberman, Judith E. / The AM5600 ambulatory blood pressure recording system. In: Blood Pressure Monitoring. 1997 ; Vol. 2, No. 3-4. pp. 193-195.
@article{7ec0bb3b9d1e4d5396e3eb12a2b4286e,
title = "The AM5600 ambulatory blood pressure recording system",
abstract = "Objective. To describe our experience with the AM5600 (Advanced Medical Products, Inc.) ambulatory blood pressure recorder-Holter monitor which has the unique features of providing full disclosure Hotter data and the storage of Korotkoff signals for each blood pressure measurement. Method. We performed a total of 236 recording sessions with 118 normotensive subjects aged 49-74 years. We then selected 25 of the recordings randomly and examined them on two different occasions, first with the aid of the Korotkoff signals and later using the traditional technique, without the aid of the Korotkoff signals. Result. Two hundred and sixteen (91.5{\%}) recordings were completed successfully. Within these data sets, on average 82.3 ± 12.1{\%} of the measurements per recording were considered acceptable after examination of the Korotkoff signals. This comparison between the traditional and the Korotkoff technique resulted in a sensitivity 11.3{\%} and a specificity 99.2{\%} of those of the traditional method. Conclusion. The performance of the AM5600 is satisfactory and the storage of Korotkoff signals improves the detection of artifacts.",
keywords = "Ambulatory blood pressure monitoring, Auscultatory, Korotkoff signal",
author = "Richey, {Phyllis A.} and Jones, {Curtis L.} and Harshfield, {Gregory A} and Somes, {Grant W.} and Johnson, {Karen C.} and Bailey, {James E.} and Soberman, {Judith E.}",
year = "1997",
month = "1",
day = "1",
language = "English (US)",
volume = "2",
pages = "193--195",
journal = "Blood Pressure Monitoring",
issn = "1359-5237",
publisher = "Lippincott Williams and Wilkins",
number = "3-4",

}

TY - JOUR

T1 - The AM5600 ambulatory blood pressure recording system

AU - Richey, Phyllis A.

AU - Jones, Curtis L.

AU - Harshfield, Gregory A

AU - Somes, Grant W.

AU - Johnson, Karen C.

AU - Bailey, James E.

AU - Soberman, Judith E.

PY - 1997/1/1

Y1 - 1997/1/1

N2 - Objective. To describe our experience with the AM5600 (Advanced Medical Products, Inc.) ambulatory blood pressure recorder-Holter monitor which has the unique features of providing full disclosure Hotter data and the storage of Korotkoff signals for each blood pressure measurement. Method. We performed a total of 236 recording sessions with 118 normotensive subjects aged 49-74 years. We then selected 25 of the recordings randomly and examined them on two different occasions, first with the aid of the Korotkoff signals and later using the traditional technique, without the aid of the Korotkoff signals. Result. Two hundred and sixteen (91.5%) recordings were completed successfully. Within these data sets, on average 82.3 ± 12.1% of the measurements per recording were considered acceptable after examination of the Korotkoff signals. This comparison between the traditional and the Korotkoff technique resulted in a sensitivity 11.3% and a specificity 99.2% of those of the traditional method. Conclusion. The performance of the AM5600 is satisfactory and the storage of Korotkoff signals improves the detection of artifacts.

AB - Objective. To describe our experience with the AM5600 (Advanced Medical Products, Inc.) ambulatory blood pressure recorder-Holter monitor which has the unique features of providing full disclosure Hotter data and the storage of Korotkoff signals for each blood pressure measurement. Method. We performed a total of 236 recording sessions with 118 normotensive subjects aged 49-74 years. We then selected 25 of the recordings randomly and examined them on two different occasions, first with the aid of the Korotkoff signals and later using the traditional technique, without the aid of the Korotkoff signals. Result. Two hundred and sixteen (91.5%) recordings were completed successfully. Within these data sets, on average 82.3 ± 12.1% of the measurements per recording were considered acceptable after examination of the Korotkoff signals. This comparison between the traditional and the Korotkoff technique resulted in a sensitivity 11.3% and a specificity 99.2% of those of the traditional method. Conclusion. The performance of the AM5600 is satisfactory and the storage of Korotkoff signals improves the detection of artifacts.

KW - Ambulatory blood pressure monitoring

KW - Auscultatory

KW - Korotkoff signal

UR - http://www.scopus.com/inward/record.url?scp=0030656434&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030656434&partnerID=8YFLogxK

M3 - Article

VL - 2

SP - 193

EP - 195

JO - Blood Pressure Monitoring

JF - Blood Pressure Monitoring

SN - 1359-5237

IS - 3-4

ER -