Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided Radiotherapy

JianYue Jin, Fang Fang Yin, Stephen E. Tenn, Paul M. Medin, Timothy D. Solberg

Research output: Contribution to journalArticle

113 Citations (Scopus)

Abstract

The ExacTrac X-Ray 6D image-guided radiotherapy (IGRT) system will be described and its performance evaluated. The system is mainly an integration of 2 subsystems: (1) an infrared (IR)-based optical positioning system (ExacTrac) and (2) a radiographic kV x-ray imaging system (X-Ray 6D). The infrared system consists of 2 IR cameras, which are used to monitor reflective body markers placed on the patient's skin to assist in patient initial setup, and an IR reflective reference star, which is attached to the treatment couch and can assist in couch movement with spatial resolution to better than 0.3 mm. The radiographic kV devices consist of 2 oblique x-ray imagers to obtain high-quality radiographs for patient position verification and adjustment. The position verification is made by fusing the radiographs with the simulation CT images using either 3 degree-of-freedom (3D) or 6 degree-of-freedom (6D) fusion algorithms. The position adjustment is performed using the infrared system according to the verification results. The reliability of the fusion algorithm will be described based on phantom and patient studies. The results indicated that the 6D fusion method is better compared to the 3D method if there are rotational deviations between the simulation and setup positions. Recently, the system has been augmented with the capabilities for image-guided positioning of targets in motion due to respiration and for gated treatment of those targets. The infrared markers provide a respiratory signal for tracking and gating of the treatment beam, with the x-ray system providing periodic confirmation of patient position relative to the gating window throughout the duration of the gated delivery.

Original languageEnglish (US)
Pages (from-to)124-134
Number of pages11
JournalMedical Dosimetry
Volume33
Issue number2
DOIs
StatePublished - Jun 1 2008

Fingerprint

Image-Guided Radiotherapy
X-Rays
Optical Devices
Respiration
Therapeutics
Equipment and Supplies
Skin

Keywords

  • Image fusion
  • Image-guided radiation therapy
  • Respiratory gating
  • X-ray guidance

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Oncology
  • Radiology Nuclear Medicine and imaging

Cite this

Jin, J., Yin, F. F., Tenn, S. E., Medin, P. M., & Solberg, T. D. (2008). Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided Radiotherapy. Medical Dosimetry, 33(2), 124-134. https://doi.org/10.1016/j.meddos.2008.02.005

Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided Radiotherapy. / Jin, JianYue; Yin, Fang Fang; Tenn, Stephen E.; Medin, Paul M.; Solberg, Timothy D.

In: Medical Dosimetry, Vol. 33, No. 2, 01.06.2008, p. 124-134.

Research output: Contribution to journalArticle

Jin, J, Yin, FF, Tenn, SE, Medin, PM & Solberg, TD 2008, 'Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided Radiotherapy', Medical Dosimetry, vol. 33, no. 2, pp. 124-134. https://doi.org/10.1016/j.meddos.2008.02.005
Jin, JianYue ; Yin, Fang Fang ; Tenn, Stephen E. ; Medin, Paul M. ; Solberg, Timothy D. / Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided Radiotherapy. In: Medical Dosimetry. 2008 ; Vol. 33, No. 2. pp. 124-134.
@article{7d103ae7eee04c618b35de770e382a99,
title = "Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided Radiotherapy",
abstract = "The ExacTrac X-Ray 6D image-guided radiotherapy (IGRT) system will be described and its performance evaluated. The system is mainly an integration of 2 subsystems: (1) an infrared (IR)-based optical positioning system (ExacTrac) and (2) a radiographic kV x-ray imaging system (X-Ray 6D). The infrared system consists of 2 IR cameras, which are used to monitor reflective body markers placed on the patient's skin to assist in patient initial setup, and an IR reflective reference star, which is attached to the treatment couch and can assist in couch movement with spatial resolution to better than 0.3 mm. The radiographic kV devices consist of 2 oblique x-ray imagers to obtain high-quality radiographs for patient position verification and adjustment. The position verification is made by fusing the radiographs with the simulation CT images using either 3 degree-of-freedom (3D) or 6 degree-of-freedom (6D) fusion algorithms. The position adjustment is performed using the infrared system according to the verification results. The reliability of the fusion algorithm will be described based on phantom and patient studies. The results indicated that the 6D fusion method is better compared to the 3D method if there are rotational deviations between the simulation and setup positions. Recently, the system has been augmented with the capabilities for image-guided positioning of targets in motion due to respiration and for gated treatment of those targets. The infrared markers provide a respiratory signal for tracking and gating of the treatment beam, with the x-ray system providing periodic confirmation of patient position relative to the gating window throughout the duration of the gated delivery.",
keywords = "Image fusion, Image-guided radiation therapy, Respiratory gating, X-ray guidance",
author = "JianYue Jin and Yin, {Fang Fang} and Tenn, {Stephen E.} and Medin, {Paul M.} and Solberg, {Timothy D.}",
year = "2008",
month = "6",
day = "1",
doi = "10.1016/j.meddos.2008.02.005",
language = "English (US)",
volume = "33",
pages = "124--134",
journal = "Medical Dosimetry",
issn = "0958-3947",
publisher = "Elsevier Inc.",
number = "2",

}

TY - JOUR

T1 - Use of the BrainLAB ExacTrac X-Ray 6D System in Image-Guided Radiotherapy

AU - Jin, JianYue

AU - Yin, Fang Fang

AU - Tenn, Stephen E.

AU - Medin, Paul M.

AU - Solberg, Timothy D.

PY - 2008/6/1

Y1 - 2008/6/1

N2 - The ExacTrac X-Ray 6D image-guided radiotherapy (IGRT) system will be described and its performance evaluated. The system is mainly an integration of 2 subsystems: (1) an infrared (IR)-based optical positioning system (ExacTrac) and (2) a radiographic kV x-ray imaging system (X-Ray 6D). The infrared system consists of 2 IR cameras, which are used to monitor reflective body markers placed on the patient's skin to assist in patient initial setup, and an IR reflective reference star, which is attached to the treatment couch and can assist in couch movement with spatial resolution to better than 0.3 mm. The radiographic kV devices consist of 2 oblique x-ray imagers to obtain high-quality radiographs for patient position verification and adjustment. The position verification is made by fusing the radiographs with the simulation CT images using either 3 degree-of-freedom (3D) or 6 degree-of-freedom (6D) fusion algorithms. The position adjustment is performed using the infrared system according to the verification results. The reliability of the fusion algorithm will be described based on phantom and patient studies. The results indicated that the 6D fusion method is better compared to the 3D method if there are rotational deviations between the simulation and setup positions. Recently, the system has been augmented with the capabilities for image-guided positioning of targets in motion due to respiration and for gated treatment of those targets. The infrared markers provide a respiratory signal for tracking and gating of the treatment beam, with the x-ray system providing periodic confirmation of patient position relative to the gating window throughout the duration of the gated delivery.

AB - The ExacTrac X-Ray 6D image-guided radiotherapy (IGRT) system will be described and its performance evaluated. The system is mainly an integration of 2 subsystems: (1) an infrared (IR)-based optical positioning system (ExacTrac) and (2) a radiographic kV x-ray imaging system (X-Ray 6D). The infrared system consists of 2 IR cameras, which are used to monitor reflective body markers placed on the patient's skin to assist in patient initial setup, and an IR reflective reference star, which is attached to the treatment couch and can assist in couch movement with spatial resolution to better than 0.3 mm. The radiographic kV devices consist of 2 oblique x-ray imagers to obtain high-quality radiographs for patient position verification and adjustment. The position verification is made by fusing the radiographs with the simulation CT images using either 3 degree-of-freedom (3D) or 6 degree-of-freedom (6D) fusion algorithms. The position adjustment is performed using the infrared system according to the verification results. The reliability of the fusion algorithm will be described based on phantom and patient studies. The results indicated that the 6D fusion method is better compared to the 3D method if there are rotational deviations between the simulation and setup positions. Recently, the system has been augmented with the capabilities for image-guided positioning of targets in motion due to respiration and for gated treatment of those targets. The infrared markers provide a respiratory signal for tracking and gating of the treatment beam, with the x-ray system providing periodic confirmation of patient position relative to the gating window throughout the duration of the gated delivery.

KW - Image fusion

KW - Image-guided radiation therapy

KW - Respiratory gating

KW - X-ray guidance

UR - http://www.scopus.com/inward/record.url?scp=43049147254&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=43049147254&partnerID=8YFLogxK

U2 - 10.1016/j.meddos.2008.02.005

DO - 10.1016/j.meddos.2008.02.005

M3 - Article

VL - 33

SP - 124

EP - 134

JO - Medical Dosimetry

JF - Medical Dosimetry

SN - 0958-3947

IS - 2

ER -