Variability of mechanical torque-limiting devices in clinical service at a US dental school

Michael Scott McCracken, Lillian Mitchell, Rashmi Hegde, Mahendra D. Mavalli

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

Purpose: The purpose of this study was to measure the variability of torque produced by a population of mechanical torque-limiting devices in clinical service in a US dental school. The torque-limiting devices were divided into two categories according to their mode of action: toggle-type and beam wrenches. Proper action of these devices is essential for calibrated delivery of preload to implant prosthetic screws. Materials and Methods: Seventeen torque-limiting devices (35 Ncm) were obtained from graduate prosthodontic, predoctoral, and faculty practice clinics. Nine of these were toggle-type devices, and eight were beam-type wrenches. Torque from each wrench was measured using an MGT electronic torque meter. Wrenches were tested in two modes, slow (over 4 seconds) and fast (over 1 second). Results: Toggle-type torque wrenches produced a mean (± SD) torque of 38.1 ± 16.0 Ncm; beam-type wrenches produced 32.8 ± 1.1 Ncm. These results were not significantly different. When tested in fast mode (1 second), toggle-type wrenches produced 28.0 ± 9.6 Ncm; in the slow mode (4 seconds) they produced significantly more force, 36.6 ± 14.0 Ncm (p < 0.001). Beam-type wrenches produced 33.2 ± 1.1 Ncm and 32.8 ± 1.1 Ncm in fast and slow modes, respectively. Conclusions: Both types of wrenches tested were capable of producing accurate torque values; however, variability was higher in the toggle-type group. Some toggle-type torque wrenches in clinical service delivered unacceptably high torque values. It is recommended that clinicians calibrate toggle-type wrenches frequently. Torque wrenches should be activated slowly, over 4 seconds, when using a correctly calibrated toggle-type wrench.

Original languageEnglish (US)
Pages (from-to)20-24
Number of pages5
JournalJournal of Prosthodontics
Volume19
Issue number1
DOIs
StatePublished - Jan 1 2010

Fingerprint

Dental Schools
Torque
Equipment and Supplies
Prosthodontics

Keywords

  • Implant
  • Mechanical torque-limiting device
  • Preload
  • Torque wrench

ASJC Scopus subject areas

  • Dentistry(all)

Cite this

Variability of mechanical torque-limiting devices in clinical service at a US dental school. / McCracken, Michael Scott; Mitchell, Lillian; Hegde, Rashmi; Mavalli, Mahendra D.

In: Journal of Prosthodontics, Vol. 19, No. 1, 01.01.2010, p. 20-24.

Research output: Contribution to journalArticle

McCracken, Michael Scott ; Mitchell, Lillian ; Hegde, Rashmi ; Mavalli, Mahendra D. / Variability of mechanical torque-limiting devices in clinical service at a US dental school. In: Journal of Prosthodontics. 2010 ; Vol. 19, No. 1. pp. 20-24.
@article{934b6228674a4bc78dfbdef72f8d9c85,
title = "Variability of mechanical torque-limiting devices in clinical service at a US dental school",
abstract = "Purpose: The purpose of this study was to measure the variability of torque produced by a population of mechanical torque-limiting devices in clinical service in a US dental school. The torque-limiting devices were divided into two categories according to their mode of action: toggle-type and beam wrenches. Proper action of these devices is essential for calibrated delivery of preload to implant prosthetic screws. Materials and Methods: Seventeen torque-limiting devices (35 Ncm) were obtained from graduate prosthodontic, predoctoral, and faculty practice clinics. Nine of these were toggle-type devices, and eight were beam-type wrenches. Torque from each wrench was measured using an MGT electronic torque meter. Wrenches were tested in two modes, slow (over 4 seconds) and fast (over 1 second). Results: Toggle-type torque wrenches produced a mean (± SD) torque of 38.1 ± 16.0 Ncm; beam-type wrenches produced 32.8 ± 1.1 Ncm. These results were not significantly different. When tested in fast mode (1 second), toggle-type wrenches produced 28.0 ± 9.6 Ncm; in the slow mode (4 seconds) they produced significantly more force, 36.6 ± 14.0 Ncm (p < 0.001). Beam-type wrenches produced 33.2 ± 1.1 Ncm and 32.8 ± 1.1 Ncm in fast and slow modes, respectively. Conclusions: Both types of wrenches tested were capable of producing accurate torque values; however, variability was higher in the toggle-type group. Some toggle-type torque wrenches in clinical service delivered unacceptably high torque values. It is recommended that clinicians calibrate toggle-type wrenches frequently. Torque wrenches should be activated slowly, over 4 seconds, when using a correctly calibrated toggle-type wrench.",
keywords = "Implant, Mechanical torque-limiting device, Preload, Torque wrench",
author = "McCracken, {Michael Scott} and Lillian Mitchell and Rashmi Hegde and Mavalli, {Mahendra D.}",
year = "2010",
month = "1",
day = "1",
doi = "10.1111/j.1532-849X.2009.00524.x",
language = "English (US)",
volume = "19",
pages = "20--24",
journal = "Journal of Prosthodontics",
issn = "1059-941X",
publisher = "Wiley-Blackwell",
number = "1",

}

TY - JOUR

T1 - Variability of mechanical torque-limiting devices in clinical service at a US dental school

AU - McCracken, Michael Scott

AU - Mitchell, Lillian

AU - Hegde, Rashmi

AU - Mavalli, Mahendra D.

PY - 2010/1/1

Y1 - 2010/1/1

N2 - Purpose: The purpose of this study was to measure the variability of torque produced by a population of mechanical torque-limiting devices in clinical service in a US dental school. The torque-limiting devices were divided into two categories according to their mode of action: toggle-type and beam wrenches. Proper action of these devices is essential for calibrated delivery of preload to implant prosthetic screws. Materials and Methods: Seventeen torque-limiting devices (35 Ncm) were obtained from graduate prosthodontic, predoctoral, and faculty practice clinics. Nine of these were toggle-type devices, and eight were beam-type wrenches. Torque from each wrench was measured using an MGT electronic torque meter. Wrenches were tested in two modes, slow (over 4 seconds) and fast (over 1 second). Results: Toggle-type torque wrenches produced a mean (± SD) torque of 38.1 ± 16.0 Ncm; beam-type wrenches produced 32.8 ± 1.1 Ncm. These results were not significantly different. When tested in fast mode (1 second), toggle-type wrenches produced 28.0 ± 9.6 Ncm; in the slow mode (4 seconds) they produced significantly more force, 36.6 ± 14.0 Ncm (p < 0.001). Beam-type wrenches produced 33.2 ± 1.1 Ncm and 32.8 ± 1.1 Ncm in fast and slow modes, respectively. Conclusions: Both types of wrenches tested were capable of producing accurate torque values; however, variability was higher in the toggle-type group. Some toggle-type torque wrenches in clinical service delivered unacceptably high torque values. It is recommended that clinicians calibrate toggle-type wrenches frequently. Torque wrenches should be activated slowly, over 4 seconds, when using a correctly calibrated toggle-type wrench.

AB - Purpose: The purpose of this study was to measure the variability of torque produced by a population of mechanical torque-limiting devices in clinical service in a US dental school. The torque-limiting devices were divided into two categories according to their mode of action: toggle-type and beam wrenches. Proper action of these devices is essential for calibrated delivery of preload to implant prosthetic screws. Materials and Methods: Seventeen torque-limiting devices (35 Ncm) were obtained from graduate prosthodontic, predoctoral, and faculty practice clinics. Nine of these were toggle-type devices, and eight were beam-type wrenches. Torque from each wrench was measured using an MGT electronic torque meter. Wrenches were tested in two modes, slow (over 4 seconds) and fast (over 1 second). Results: Toggle-type torque wrenches produced a mean (± SD) torque of 38.1 ± 16.0 Ncm; beam-type wrenches produced 32.8 ± 1.1 Ncm. These results were not significantly different. When tested in fast mode (1 second), toggle-type wrenches produced 28.0 ± 9.6 Ncm; in the slow mode (4 seconds) they produced significantly more force, 36.6 ± 14.0 Ncm (p < 0.001). Beam-type wrenches produced 33.2 ± 1.1 Ncm and 32.8 ± 1.1 Ncm in fast and slow modes, respectively. Conclusions: Both types of wrenches tested were capable of producing accurate torque values; however, variability was higher in the toggle-type group. Some toggle-type torque wrenches in clinical service delivered unacceptably high torque values. It is recommended that clinicians calibrate toggle-type wrenches frequently. Torque wrenches should be activated slowly, over 4 seconds, when using a correctly calibrated toggle-type wrench.

KW - Implant

KW - Mechanical torque-limiting device

KW - Preload

KW - Torque wrench

UR - http://www.scopus.com/inward/record.url?scp=76649144483&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=76649144483&partnerID=8YFLogxK

U2 - 10.1111/j.1532-849X.2009.00524.x

DO - 10.1111/j.1532-849X.2009.00524.x

M3 - Article

C2 - 19765196

AN - SCOPUS:76649144483

VL - 19

SP - 20

EP - 24

JO - Journal of Prosthodontics

JF - Journal of Prosthodontics

SN - 1059-941X

IS - 1

ER -