WE‐E‐BRB‐04: Quantitative Dose Tracking Enabled Through a Novel Deformable 3D Dosimeter

T. Juang, J. Newton, S. Das, J. Adamovics, M. Oldham

Research output: Contribution to journalArticle

Abstract

Purpose: To evaluate and investigate the feasibility of a new method for validating dose tracking algorithms in deforming tissues using a novel deformable 3D dosimeter. Methods: A novel deformable 3D Presage dosimeter is reported consisting of a stretchy polyurethane matrix doped with radiochromic leuco‐dye. Two deformable cylindrical dosimeters (6 cm diameter, 5 cm long) were manufactured and irradiated with a checkerboard arrangement of 5 mm square pencil beams created by MLC fields. One dosimeter was irradiated under lateral compression by 33% (6 cm down to 4 cm diameter) to simulate a deformed organ. A second control dosimeter was irradiated with the same checkerboard pattern but without deformation applied. High‐resolution 3D dose distributions (isotropic 1 mm resolution) were obtained by optical‐CT imaging. Physical dose deformation was quantified by comparing checkerboard pencil beam shapes and positions in the deformed and control dosimeters. Results: Deformation of dose in the deformed dosimeter was clearly visible in all 3 dimensions. The deformed checkerboard dose pattern showed expansion of 16% – 46% along the axis of compression, with higher expansion observed in the central regions of the dosimeter. Perpendicular to the compression axis, the dose pattern contracted by 7% – 13%. Peak dose changes of −6% and +30% were observed parallel and perpendicular to the compression axis respectively. Dose response was linear from 0–8 Gy. Conclusions: Dose tracking was successfully quantified in a novel deforming 3D dosimeter. This capability has potential as a powerful new method for validating deformable dose tracking and registration algorithms.

Original languageEnglish (US)
Number of pages1
JournalMedical Physics
Volume39
Issue number6
DOIs
StatePublished - Jun 2012

Fingerprint

Radiation Dosimeters
Polyurethanes

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

WE‐E‐BRB‐04 : Quantitative Dose Tracking Enabled Through a Novel Deformable 3D Dosimeter. / Juang, T.; Newton, J.; Das, S.; Adamovics, J.; Oldham, M.

In: Medical Physics, Vol. 39, No. 6, 06.2012.

Research output: Contribution to journalArticle

@article{6ee445f64c0e4cce9426b85cc09424b0,
title = "WE‐E‐BRB‐04: Quantitative Dose Tracking Enabled Through a Novel Deformable 3D Dosimeter",
abstract = "Purpose: To evaluate and investigate the feasibility of a new method for validating dose tracking algorithms in deforming tissues using a novel deformable 3D dosimeter. Methods: A novel deformable 3D Presage dosimeter is reported consisting of a stretchy polyurethane matrix doped with radiochromic leuco‐dye. Two deformable cylindrical dosimeters (6 cm diameter, 5 cm long) were manufactured and irradiated with a checkerboard arrangement of 5 mm square pencil beams created by MLC fields. One dosimeter was irradiated under lateral compression by 33{\%} (6 cm down to 4 cm diameter) to simulate a deformed organ. A second control dosimeter was irradiated with the same checkerboard pattern but without deformation applied. High‐resolution 3D dose distributions (isotropic 1 mm resolution) were obtained by optical‐CT imaging. Physical dose deformation was quantified by comparing checkerboard pencil beam shapes and positions in the deformed and control dosimeters. Results: Deformation of dose in the deformed dosimeter was clearly visible in all 3 dimensions. The deformed checkerboard dose pattern showed expansion of 16{\%} – 46{\%} along the axis of compression, with higher expansion observed in the central regions of the dosimeter. Perpendicular to the compression axis, the dose pattern contracted by 7{\%} – 13{\%}. Peak dose changes of −6{\%} and +30{\%} were observed parallel and perpendicular to the compression axis respectively. Dose response was linear from 0–8 Gy. Conclusions: Dose tracking was successfully quantified in a novel deforming 3D dosimeter. This capability has potential as a powerful new method for validating deformable dose tracking and registration algorithms.",
author = "T. Juang and J. Newton and S. Das and J. Adamovics and M. Oldham",
year = "2012",
month = "6",
doi = "10.1118/1.4736146",
language = "English (US)",
volume = "39",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - WE‐E‐BRB‐04

T2 - Quantitative Dose Tracking Enabled Through a Novel Deformable 3D Dosimeter

AU - Juang, T.

AU - Newton, J.

AU - Das, S.

AU - Adamovics, J.

AU - Oldham, M.

PY - 2012/6

Y1 - 2012/6

N2 - Purpose: To evaluate and investigate the feasibility of a new method for validating dose tracking algorithms in deforming tissues using a novel deformable 3D dosimeter. Methods: A novel deformable 3D Presage dosimeter is reported consisting of a stretchy polyurethane matrix doped with radiochromic leuco‐dye. Two deformable cylindrical dosimeters (6 cm diameter, 5 cm long) were manufactured and irradiated with a checkerboard arrangement of 5 mm square pencil beams created by MLC fields. One dosimeter was irradiated under lateral compression by 33% (6 cm down to 4 cm diameter) to simulate a deformed organ. A second control dosimeter was irradiated with the same checkerboard pattern but without deformation applied. High‐resolution 3D dose distributions (isotropic 1 mm resolution) were obtained by optical‐CT imaging. Physical dose deformation was quantified by comparing checkerboard pencil beam shapes and positions in the deformed and control dosimeters. Results: Deformation of dose in the deformed dosimeter was clearly visible in all 3 dimensions. The deformed checkerboard dose pattern showed expansion of 16% – 46% along the axis of compression, with higher expansion observed in the central regions of the dosimeter. Perpendicular to the compression axis, the dose pattern contracted by 7% – 13%. Peak dose changes of −6% and +30% were observed parallel and perpendicular to the compression axis respectively. Dose response was linear from 0–8 Gy. Conclusions: Dose tracking was successfully quantified in a novel deforming 3D dosimeter. This capability has potential as a powerful new method for validating deformable dose tracking and registration algorithms.

AB - Purpose: To evaluate and investigate the feasibility of a new method for validating dose tracking algorithms in deforming tissues using a novel deformable 3D dosimeter. Methods: A novel deformable 3D Presage dosimeter is reported consisting of a stretchy polyurethane matrix doped with radiochromic leuco‐dye. Two deformable cylindrical dosimeters (6 cm diameter, 5 cm long) were manufactured and irradiated with a checkerboard arrangement of 5 mm square pencil beams created by MLC fields. One dosimeter was irradiated under lateral compression by 33% (6 cm down to 4 cm diameter) to simulate a deformed organ. A second control dosimeter was irradiated with the same checkerboard pattern but without deformation applied. High‐resolution 3D dose distributions (isotropic 1 mm resolution) were obtained by optical‐CT imaging. Physical dose deformation was quantified by comparing checkerboard pencil beam shapes and positions in the deformed and control dosimeters. Results: Deformation of dose in the deformed dosimeter was clearly visible in all 3 dimensions. The deformed checkerboard dose pattern showed expansion of 16% – 46% along the axis of compression, with higher expansion observed in the central regions of the dosimeter. Perpendicular to the compression axis, the dose pattern contracted by 7% – 13%. Peak dose changes of −6% and +30% were observed parallel and perpendicular to the compression axis respectively. Dose response was linear from 0–8 Gy. Conclusions: Dose tracking was successfully quantified in a novel deforming 3D dosimeter. This capability has potential as a powerful new method for validating deformable dose tracking and registration algorithms.

UR - http://www.scopus.com/inward/record.url?scp=85024776856&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85024776856&partnerID=8YFLogxK

U2 - 10.1118/1.4736146

DO - 10.1118/1.4736146

M3 - Article

AN - SCOPUS:85024776856

VL - 39

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -