Avoidance of electrode related MRI artifact during staged deep brain stimulator implantation

Cole A. Giller, Shyamal Mehta, Nathan Eugene Yanasak, Patrick Jenkins

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Background Centers implanting deep brain stimulator (DBS) electrodes on different days often protect the first electrode tip with a protective cap, tunnel it under the scalp, and connect it to the generator at a later procedure. If magnetic resonance imaging (MRI) is used for planning during the second implantation, MRI artifacts from the protective cap could potentially corrupt the stereotactic coordinates. The importance of this problem may increase if emerging MRI safety data lead to more frequent use of MRI for these purposes. Objective To describe an MRI artifact arising from the use of the standard protective DBS cap that corrupts stereotactic planning and to describe a way to avoid the artifact. Methods After noting the artifact during a staged DBS procedure, a nonmetallic silastic sleeve contained in the existing DBS implantation kit was used in nine subsequent patients. Two caps with standard metallic screws were also tested with MRI phantoms. Results The silastic sleeve protected the DBS electrode but did not produce MRI artifact. The phantom studies demonstrated significant artifact from caps containing screws. Conclusion A silastic sleeve provides adequate protection of the DBS electrode during staged implantation and avoids the MRI artifact associated with protective caps with screws.

Original languageEnglish (US)
Pages (from-to)320-323
Number of pages4
JournalJournal of Neurological Surgery, Part A: Central European Neurosurgery
Volume73
Issue number5
DOIs
StatePublished - Sep 1 2012

Fingerprint

Artifacts
Electrodes
Magnetic Resonance Imaging
Brain
Imaging Phantoms
Scalp
Safety
baysilon

Keywords

  • Deep brain stimulation
  • MRI artifact
  • Parkinson's disease

ASJC Scopus subject areas

  • Surgery
  • Clinical Neurology

Cite this

Avoidance of electrode related MRI artifact during staged deep brain stimulator implantation. / Giller, Cole A.; Mehta, Shyamal; Yanasak, Nathan Eugene; Jenkins, Patrick.

In: Journal of Neurological Surgery, Part A: Central European Neurosurgery, Vol. 73, No. 5, 01.09.2012, p. 320-323.

Research output: Contribution to journalArticle

@article{e1a2618bd0eb45b4af99835169f36d3a,
title = "Avoidance of electrode related MRI artifact during staged deep brain stimulator implantation",
abstract = "Background Centers implanting deep brain stimulator (DBS) electrodes on different days often protect the first electrode tip with a protective cap, tunnel it under the scalp, and connect it to the generator at a later procedure. If magnetic resonance imaging (MRI) is used for planning during the second implantation, MRI artifacts from the protective cap could potentially corrupt the stereotactic coordinates. The importance of this problem may increase if emerging MRI safety data lead to more frequent use of MRI for these purposes. Objective To describe an MRI artifact arising from the use of the standard protective DBS cap that corrupts stereotactic planning and to describe a way to avoid the artifact. Methods After noting the artifact during a staged DBS procedure, a nonmetallic silastic sleeve contained in the existing DBS implantation kit was used in nine subsequent patients. Two caps with standard metallic screws were also tested with MRI phantoms. Results The silastic sleeve protected the DBS electrode but did not produce MRI artifact. The phantom studies demonstrated significant artifact from caps containing screws. Conclusion A silastic sleeve provides adequate protection of the DBS electrode during staged implantation and avoids the MRI artifact associated with protective caps with screws.",
keywords = "Deep brain stimulation, MRI artifact, Parkinson's disease",
author = "Giller, {Cole A.} and Shyamal Mehta and Yanasak, {Nathan Eugene} and Patrick Jenkins",
year = "2012",
month = "9",
day = "1",
doi = "10.1055/s-0032-1322590",
language = "English (US)",
volume = "73",
pages = "320--323",
journal = "Journal of Neurological Surgery, Part A: Central European Neurosurgery",
issn = "2193-6315",
publisher = "Thieme",
number = "5",

}

TY - JOUR

T1 - Avoidance of electrode related MRI artifact during staged deep brain stimulator implantation

AU - Giller, Cole A.

AU - Mehta, Shyamal

AU - Yanasak, Nathan Eugene

AU - Jenkins, Patrick

PY - 2012/9/1

Y1 - 2012/9/1

N2 - Background Centers implanting deep brain stimulator (DBS) electrodes on different days often protect the first electrode tip with a protective cap, tunnel it under the scalp, and connect it to the generator at a later procedure. If magnetic resonance imaging (MRI) is used for planning during the second implantation, MRI artifacts from the protective cap could potentially corrupt the stereotactic coordinates. The importance of this problem may increase if emerging MRI safety data lead to more frequent use of MRI for these purposes. Objective To describe an MRI artifact arising from the use of the standard protective DBS cap that corrupts stereotactic planning and to describe a way to avoid the artifact. Methods After noting the artifact during a staged DBS procedure, a nonmetallic silastic sleeve contained in the existing DBS implantation kit was used in nine subsequent patients. Two caps with standard metallic screws were also tested with MRI phantoms. Results The silastic sleeve protected the DBS electrode but did not produce MRI artifact. The phantom studies demonstrated significant artifact from caps containing screws. Conclusion A silastic sleeve provides adequate protection of the DBS electrode during staged implantation and avoids the MRI artifact associated with protective caps with screws.

AB - Background Centers implanting deep brain stimulator (DBS) electrodes on different days often protect the first electrode tip with a protective cap, tunnel it under the scalp, and connect it to the generator at a later procedure. If magnetic resonance imaging (MRI) is used for planning during the second implantation, MRI artifacts from the protective cap could potentially corrupt the stereotactic coordinates. The importance of this problem may increase if emerging MRI safety data lead to more frequent use of MRI for these purposes. Objective To describe an MRI artifact arising from the use of the standard protective DBS cap that corrupts stereotactic planning and to describe a way to avoid the artifact. Methods After noting the artifact during a staged DBS procedure, a nonmetallic silastic sleeve contained in the existing DBS implantation kit was used in nine subsequent patients. Two caps with standard metallic screws were also tested with MRI phantoms. Results The silastic sleeve protected the DBS electrode but did not produce MRI artifact. The phantom studies demonstrated significant artifact from caps containing screws. Conclusion A silastic sleeve provides adequate protection of the DBS electrode during staged implantation and avoids the MRI artifact associated with protective caps with screws.

KW - Deep brain stimulation

KW - MRI artifact

KW - Parkinson's disease

UR - http://www.scopus.com/inward/record.url?scp=84873270611&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84873270611&partnerID=8YFLogxK

U2 - 10.1055/s-0032-1322590

DO - 10.1055/s-0032-1322590

M3 - Article

C2 - 22847067

AN - SCOPUS:84873270611

VL - 73

SP - 320

EP - 323

JO - Journal of Neurological Surgery, Part A: Central European Neurosurgery

JF - Journal of Neurological Surgery, Part A: Central European Neurosurgery

SN - 2193-6315

IS - 5

ER -