Continuous spectrophotometric measurements of arteriovenous oxygen difference

A. P. Shepherd, John C. Sutherland, A. F. Wilson

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

A relatively inexpensive dual beam optical absorbance monitor was tested and adapted to measure arteriovenous oxygen difference ((a-v)ΔO2). Since the molar extinction coefficients of oxy and deoxyhemoglobin are markedly different at 660 nm, the difference in optical density between arterial and venous blood, as measured by the absorbance monitor, was shown to be linearly related to (a-v)ΔO2. When appropriate cuvettes (optical path length = 0.5 mm) were used, the instrument's 90% response time (4.4 sec) was sufficiently rapid for most physical applications such as Fick determinations of the rate of an organ's oxygen consumption. The theoretical basis of this spectrophotometric (a-v)ΔO2 measurement and sources of error are discussed in detail.

Original languageEnglish (US)
Pages (from-to)152-155
Number of pages4
JournalJournal of Applied Physiology
Volume39
Issue number1
DOIs
StatePublished - Jan 1 1975
Externally publishedYes

Fingerprint

Oxygen Consumption
Reaction Time
Research Design
Oxygen
Psychological Extinction
deoxyhemoglobin

ASJC Scopus subject areas

  • Physiology
  • Physiology (medical)

Cite this

Continuous spectrophotometric measurements of arteriovenous oxygen difference. / Shepherd, A. P.; Sutherland, John C.; Wilson, A. F.

In: Journal of Applied Physiology, Vol. 39, No. 1, 01.01.1975, p. 152-155.

Research output: Contribution to journalArticle

@article{7f8388229565490499bebfe530898522,
title = "Continuous spectrophotometric measurements of arteriovenous oxygen difference",
abstract = "A relatively inexpensive dual beam optical absorbance monitor was tested and adapted to measure arteriovenous oxygen difference ((a-v)ΔO2). Since the molar extinction coefficients of oxy and deoxyhemoglobin are markedly different at 660 nm, the difference in optical density between arterial and venous blood, as measured by the absorbance monitor, was shown to be linearly related to (a-v)ΔO2. When appropriate cuvettes (optical path length = 0.5 mm) were used, the instrument's 90{\%} response time (4.4 sec) was sufficiently rapid for most physical applications such as Fick determinations of the rate of an organ's oxygen consumption. The theoretical basis of this spectrophotometric (a-v)ΔO2 measurement and sources of error are discussed in detail.",
author = "Shepherd, {A. P.} and Sutherland, {John C.} and Wilson, {A. F.}",
year = "1975",
month = "1",
day = "1",
doi = "10.1152/jappl.1975.39.1.152",
language = "English (US)",
volume = "39",
pages = "152--155",
journal = "Journal of Applied Physiology Respiratory Environmental and Exercise Physiology",
issn = "8750-7587",
publisher = "American Physiological Society",
number = "1",

}

TY - JOUR

T1 - Continuous spectrophotometric measurements of arteriovenous oxygen difference

AU - Shepherd, A. P.

AU - Sutherland, John C.

AU - Wilson, A. F.

PY - 1975/1/1

Y1 - 1975/1/1

N2 - A relatively inexpensive dual beam optical absorbance monitor was tested and adapted to measure arteriovenous oxygen difference ((a-v)ΔO2). Since the molar extinction coefficients of oxy and deoxyhemoglobin are markedly different at 660 nm, the difference in optical density between arterial and venous blood, as measured by the absorbance monitor, was shown to be linearly related to (a-v)ΔO2. When appropriate cuvettes (optical path length = 0.5 mm) were used, the instrument's 90% response time (4.4 sec) was sufficiently rapid for most physical applications such as Fick determinations of the rate of an organ's oxygen consumption. The theoretical basis of this spectrophotometric (a-v)ΔO2 measurement and sources of error are discussed in detail.

AB - A relatively inexpensive dual beam optical absorbance monitor was tested and adapted to measure arteriovenous oxygen difference ((a-v)ΔO2). Since the molar extinction coefficients of oxy and deoxyhemoglobin are markedly different at 660 nm, the difference in optical density between arterial and venous blood, as measured by the absorbance monitor, was shown to be linearly related to (a-v)ΔO2. When appropriate cuvettes (optical path length = 0.5 mm) were used, the instrument's 90% response time (4.4 sec) was sufficiently rapid for most physical applications such as Fick determinations of the rate of an organ's oxygen consumption. The theoretical basis of this spectrophotometric (a-v)ΔO2 measurement and sources of error are discussed in detail.

UR - http://www.scopus.com/inward/record.url?scp=0016820842&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0016820842&partnerID=8YFLogxK

U2 - 10.1152/jappl.1975.39.1.152

DO - 10.1152/jappl.1975.39.1.152

M3 - Article

C2 - 1150582

AN - SCOPUS:0016820842

VL - 39

SP - 152

EP - 155

JO - Journal of Applied Physiology Respiratory Environmental and Exercise Physiology

JF - Journal of Applied Physiology Respiratory Environmental and Exercise Physiology

SN - 8750-7587

IS - 1

ER -