Defibrillation energy requirements and electrical heterogeneity during total body hypothermia

M. R. Ujhelyi, J. J. Sims, S. A. Dubin, John R Vender, A. W. Miller

Research output: Contribution to journalArticle

36 Citations (Scopus)

Abstract

Objective: Determine the effects of hypothermia on defibrillation energy requirements and cardiac electrophysiology. Design: Prospective randomized acute intervention trial. Setting: Medical center animal laboratory. Subjects: Fifteen domestic farm swine. Interventions: Swine were randomized to a hypothermia group (n = 8) or a control group (n = 7). All animals were instrumented with a transvenous defibrillation system connected to a defibrillator that delivers a biphasic-truncated waveform. Values for defibrillation energy requirements were measured at baseline (normothermia, 38-40°C) and during treatment with total body hypothermia (30°C) or no temperature change (sham). Hypothermia was induced by circulating ice-water through anterior and posterior surgical thermal blankets. Measurements and Main Results: Defibrillation energy requirement values at 20%, 50%, and 80% were determined by using an up/down method. In the hypothermia group, defibrillation energy requirement values at baseline did not significantly change during hypothermia (defibrillation energy requirements 50% = 14 ± 2 J vs. 15 ± 2 J, respectively). Similarly, the defibrillation energy requirement values in the control group did not change from baseline to sham phase (defibrillation energy requirements 50% = 12 ± 1 J vs. 13 ± 1 J, respectively). Hypothermia profoundly affected cardiac electrophysiology, decreasing ventricular fibrillation threshold by 72%, conduction velocity by 25% (p < .01), and tissue excitability, while it prolonged ventricular repolarization and refractoriness by 7.5% to 15%, respectively (p < .05). Conclusions: Total body cooling to 30°C was highly arrhythmogenic, although this unstable electrophysiological state did not alter ventricular defibrillation energy requirements. These data suggest that hypothermia may be used to slow metabolic processes without concern over the ability to successfully defibrillate and treat hypothermia-induced arrhythmias.

Original languageEnglish (US)
Pages (from-to)1006-1011
Number of pages6
JournalCritical Care Medicine
Volume29
Issue number5
DOIs
StatePublished - Jan 1 2001

Fingerprint

Hypothermia
Cardiac Electrophysiology
Induced Hypothermia
Swine
Control Groups
Defibrillators
Laboratory Animals
Ice
Ventricular Fibrillation
Cardiac Arrhythmias
Hot Temperature
Temperature
Water

Keywords

  • Conduction velocity
  • Electrical heterogeneity
  • Electrical stimulation
  • Electrophysiology
  • Hypothermia
  • Ion conductance
  • Refractoriness
  • Ventricular defibrillation
  • Ventricular fibrillation

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine

Cite this

Defibrillation energy requirements and electrical heterogeneity during total body hypothermia. / Ujhelyi, M. R.; Sims, J. J.; Dubin, S. A.; Vender, John R; Miller, A. W.

In: Critical Care Medicine, Vol. 29, No. 5, 01.01.2001, p. 1006-1011.

Research output: Contribution to journalArticle

Ujhelyi, M. R. ; Sims, J. J. ; Dubin, S. A. ; Vender, John R ; Miller, A. W. / Defibrillation energy requirements and electrical heterogeneity during total body hypothermia. In: Critical Care Medicine. 2001 ; Vol. 29, No. 5. pp. 1006-1011.
@article{dbf4a36f91704949a4d8f52b3e5976f8,
title = "Defibrillation energy requirements and electrical heterogeneity during total body hypothermia",
abstract = "Objective: Determine the effects of hypothermia on defibrillation energy requirements and cardiac electrophysiology. Design: Prospective randomized acute intervention trial. Setting: Medical center animal laboratory. Subjects: Fifteen domestic farm swine. Interventions: Swine were randomized to a hypothermia group (n = 8) or a control group (n = 7). All animals were instrumented with a transvenous defibrillation system connected to a defibrillator that delivers a biphasic-truncated waveform. Values for defibrillation energy requirements were measured at baseline (normothermia, 38-40°C) and during treatment with total body hypothermia (30°C) or no temperature change (sham). Hypothermia was induced by circulating ice-water through anterior and posterior surgical thermal blankets. Measurements and Main Results: Defibrillation energy requirement values at 20{\%}, 50{\%}, and 80{\%} were determined by using an up/down method. In the hypothermia group, defibrillation energy requirement values at baseline did not significantly change during hypothermia (defibrillation energy requirements 50{\%} = 14 ± 2 J vs. 15 ± 2 J, respectively). Similarly, the defibrillation energy requirement values in the control group did not change from baseline to sham phase (defibrillation energy requirements 50{\%} = 12 ± 1 J vs. 13 ± 1 J, respectively). Hypothermia profoundly affected cardiac electrophysiology, decreasing ventricular fibrillation threshold by 72{\%}, conduction velocity by 25{\%} (p < .01), and tissue excitability, while it prolonged ventricular repolarization and refractoriness by 7.5{\%} to 15{\%}, respectively (p < .05). Conclusions: Total body cooling to 30°C was highly arrhythmogenic, although this unstable electrophysiological state did not alter ventricular defibrillation energy requirements. These data suggest that hypothermia may be used to slow metabolic processes without concern over the ability to successfully defibrillate and treat hypothermia-induced arrhythmias.",
keywords = "Conduction velocity, Electrical heterogeneity, Electrical stimulation, Electrophysiology, Hypothermia, Ion conductance, Refractoriness, Ventricular defibrillation, Ventricular fibrillation",
author = "Ujhelyi, {M. R.} and Sims, {J. J.} and Dubin, {S. A.} and Vender, {John R} and Miller, {A. W.}",
year = "2001",
month = "1",
day = "1",
doi = "10.1097/00003246-200105000-00025",
language = "English (US)",
volume = "29",
pages = "1006--1011",
journal = "Critical Care Medicine",
issn = "0090-3493",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - Defibrillation energy requirements and electrical heterogeneity during total body hypothermia

AU - Ujhelyi, M. R.

AU - Sims, J. J.

AU - Dubin, S. A.

AU - Vender, John R

AU - Miller, A. W.

PY - 2001/1/1

Y1 - 2001/1/1

N2 - Objective: Determine the effects of hypothermia on defibrillation energy requirements and cardiac electrophysiology. Design: Prospective randomized acute intervention trial. Setting: Medical center animal laboratory. Subjects: Fifteen domestic farm swine. Interventions: Swine were randomized to a hypothermia group (n = 8) or a control group (n = 7). All animals were instrumented with a transvenous defibrillation system connected to a defibrillator that delivers a biphasic-truncated waveform. Values for defibrillation energy requirements were measured at baseline (normothermia, 38-40°C) and during treatment with total body hypothermia (30°C) or no temperature change (sham). Hypothermia was induced by circulating ice-water through anterior and posterior surgical thermal blankets. Measurements and Main Results: Defibrillation energy requirement values at 20%, 50%, and 80% were determined by using an up/down method. In the hypothermia group, defibrillation energy requirement values at baseline did not significantly change during hypothermia (defibrillation energy requirements 50% = 14 ± 2 J vs. 15 ± 2 J, respectively). Similarly, the defibrillation energy requirement values in the control group did not change from baseline to sham phase (defibrillation energy requirements 50% = 12 ± 1 J vs. 13 ± 1 J, respectively). Hypothermia profoundly affected cardiac electrophysiology, decreasing ventricular fibrillation threshold by 72%, conduction velocity by 25% (p < .01), and tissue excitability, while it prolonged ventricular repolarization and refractoriness by 7.5% to 15%, respectively (p < .05). Conclusions: Total body cooling to 30°C was highly arrhythmogenic, although this unstable electrophysiological state did not alter ventricular defibrillation energy requirements. These data suggest that hypothermia may be used to slow metabolic processes without concern over the ability to successfully defibrillate and treat hypothermia-induced arrhythmias.

AB - Objective: Determine the effects of hypothermia on defibrillation energy requirements and cardiac electrophysiology. Design: Prospective randomized acute intervention trial. Setting: Medical center animal laboratory. Subjects: Fifteen domestic farm swine. Interventions: Swine were randomized to a hypothermia group (n = 8) or a control group (n = 7). All animals were instrumented with a transvenous defibrillation system connected to a defibrillator that delivers a biphasic-truncated waveform. Values for defibrillation energy requirements were measured at baseline (normothermia, 38-40°C) and during treatment with total body hypothermia (30°C) or no temperature change (sham). Hypothermia was induced by circulating ice-water through anterior and posterior surgical thermal blankets. Measurements and Main Results: Defibrillation energy requirement values at 20%, 50%, and 80% were determined by using an up/down method. In the hypothermia group, defibrillation energy requirement values at baseline did not significantly change during hypothermia (defibrillation energy requirements 50% = 14 ± 2 J vs. 15 ± 2 J, respectively). Similarly, the defibrillation energy requirement values in the control group did not change from baseline to sham phase (defibrillation energy requirements 50% = 12 ± 1 J vs. 13 ± 1 J, respectively). Hypothermia profoundly affected cardiac electrophysiology, decreasing ventricular fibrillation threshold by 72%, conduction velocity by 25% (p < .01), and tissue excitability, while it prolonged ventricular repolarization and refractoriness by 7.5% to 15%, respectively (p < .05). Conclusions: Total body cooling to 30°C was highly arrhythmogenic, although this unstable electrophysiological state did not alter ventricular defibrillation energy requirements. These data suggest that hypothermia may be used to slow metabolic processes without concern over the ability to successfully defibrillate and treat hypothermia-induced arrhythmias.

KW - Conduction velocity

KW - Electrical heterogeneity

KW - Electrical stimulation

KW - Electrophysiology

KW - Hypothermia

KW - Ion conductance

KW - Refractoriness

KW - Ventricular defibrillation

KW - Ventricular fibrillation

UR - http://www.scopus.com/inward/record.url?scp=0035011544&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035011544&partnerID=8YFLogxK

U2 - 10.1097/00003246-200105000-00025

DO - 10.1097/00003246-200105000-00025

M3 - Article

VL - 29

SP - 1006

EP - 1011

JO - Critical Care Medicine

JF - Critical Care Medicine

SN - 0090-3493

IS - 5

ER -