Minimizing interference and false alarms from electrocautery in the Nellcor N-100 pulse oximeter

Frank E Block, George J. Detko

Research output: Contribution to journalArticle

11 Citations (Scopus)

Abstract

Because the Nellcor N-100 pulse oximeter is generally quite sensitive to electrocautery interference, erroneous oxygen saturation values may be displayed and an audible false alarm may sound. We have discovered, through trial and error, several techniques that may minimize both the interference and the false alarms. Keeping the pulse oximeter sensor and the oximeter unit as far as possible from the surgical site and the electrocautery grounding plate will usually decrease the interference. If this step is inadequate, the incidence of false alarms can be decreased by setting the high pulse rate alarm to approximately 160, operating the unit in mode 2 (the rapid-response mode), and setting the "alarm off" function to 120 seconds. These steps will usually markedly decrease the incidence of false alarms.

Original languageEnglish (US)
Pages (from-to)203-205
Number of pages3
JournalJournal of Clinical Monitoring
Volume2
Issue number3
DOIs
StatePublished - Jul 1 1986
Externally publishedYes

Fingerprint

Electrocoagulation
Incidence
Heart Rate
Oxygen

Keywords

  • alarms
  • artifact
  • Electrical systems
  • electrocautery
  • Equipment
  • interference
  • oximeters

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine

Cite this

Minimizing interference and false alarms from electrocautery in the Nellcor N-100 pulse oximeter. / Block, Frank E; Detko, George J.

In: Journal of Clinical Monitoring, Vol. 2, No. 3, 01.07.1986, p. 203-205.

Research output: Contribution to journalArticle

@article{5b08c92123224ebebf1ad0db45f8bb31,
title = "Minimizing interference and false alarms from electrocautery in the Nellcor N-100 pulse oximeter",
abstract = "Because the Nellcor N-100 pulse oximeter is generally quite sensitive to electrocautery interference, erroneous oxygen saturation values may be displayed and an audible false alarm may sound. We have discovered, through trial and error, several techniques that may minimize both the interference and the false alarms. Keeping the pulse oximeter sensor and the oximeter unit as far as possible from the surgical site and the electrocautery grounding plate will usually decrease the interference. If this step is inadequate, the incidence of false alarms can be decreased by setting the high pulse rate alarm to approximately 160, operating the unit in mode 2 (the rapid-response mode), and setting the {"}alarm off{"} function to 120 seconds. These steps will usually markedly decrease the incidence of false alarms.",
keywords = "alarms, artifact, Electrical systems, electrocautery, Equipment, interference, oximeters",
author = "Block, {Frank E} and Detko, {George J.}",
year = "1986",
month = "7",
day = "1",
doi = "10.1007/BF01620553",
language = "English (US)",
volume = "2",
pages = "203--205",
journal = "Journal of Clinical Monitoring and Computing",
issn = "1387-1307",
publisher = "Springer Netherlands",
number = "3",

}

TY - JOUR

T1 - Minimizing interference and false alarms from electrocautery in the Nellcor N-100 pulse oximeter

AU - Block, Frank E

AU - Detko, George J.

PY - 1986/7/1

Y1 - 1986/7/1

N2 - Because the Nellcor N-100 pulse oximeter is generally quite sensitive to electrocautery interference, erroneous oxygen saturation values may be displayed and an audible false alarm may sound. We have discovered, through trial and error, several techniques that may minimize both the interference and the false alarms. Keeping the pulse oximeter sensor and the oximeter unit as far as possible from the surgical site and the electrocautery grounding plate will usually decrease the interference. If this step is inadequate, the incidence of false alarms can be decreased by setting the high pulse rate alarm to approximately 160, operating the unit in mode 2 (the rapid-response mode), and setting the "alarm off" function to 120 seconds. These steps will usually markedly decrease the incidence of false alarms.

AB - Because the Nellcor N-100 pulse oximeter is generally quite sensitive to electrocautery interference, erroneous oxygen saturation values may be displayed and an audible false alarm may sound. We have discovered, through trial and error, several techniques that may minimize both the interference and the false alarms. Keeping the pulse oximeter sensor and the oximeter unit as far as possible from the surgical site and the electrocautery grounding plate will usually decrease the interference. If this step is inadequate, the incidence of false alarms can be decreased by setting the high pulse rate alarm to approximately 160, operating the unit in mode 2 (the rapid-response mode), and setting the "alarm off" function to 120 seconds. These steps will usually markedly decrease the incidence of false alarms.

KW - alarms

KW - artifact

KW - Electrical systems

KW - electrocautery

KW - Equipment

KW - interference

KW - oximeters

UR - http://www.scopus.com/inward/record.url?scp=0022511724&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0022511724&partnerID=8YFLogxK

U2 - 10.1007/BF01620553

DO - 10.1007/BF01620553

M3 - Article

C2 - 3746371

AN - SCOPUS:0022511724

VL - 2

SP - 203

EP - 205

JO - Journal of Clinical Monitoring and Computing

JF - Journal of Clinical Monitoring and Computing

SN - 1387-1307

IS - 3

ER -