Performance characteristics of transmission imaging using a uniform sheet source with parallel-hole collimation

Zongjian Cao, Beniamin M W Tsui

Research output: Contribution to journalArticle

25 Citations (Scopus)

Abstract

Transmission imaging is receiving increasing attention in SPECT due to the need to compensate for nonuniform attenuation in cardiac-chest SPECT. The quality of a transmission image has an important effect on the measured attenuation distribution. To improve image quality, knowledge of the performance characteristics of a transmission imaging system is essential. The characteristics, spatial resolution, detection efficiency, photon flux, and exposure to the object, of a transmission imaging system consisting of a gamma camera and a uniform sheet source have been studied. The results demonstrate that spatial resolution of a transmission imaging system can be improved by use of a high-resolution source collimator at the price of a moderate decrease in detection efficiency, in comparison to the uncollimated case. Also, the source collimator significantly reduces the photon flux and exposure to the object. This investigation suggests that a high-resolution collimator be used with an intense sheet source to improve spatial resolution and reduce statistical noise with low exposure to the patient. This research further suggests that the amount of source activity is determined by the requirement of image quality, detection geometry, and allowed absorbed dose to the patient.

Original languageEnglish (US)
Pages (from-to)1205-1212
Number of pages8
JournalMedical Physics
Volume19
Issue number5
DOIs
StatePublished - Jan 1 1992

Fingerprint

Single-Photon Emission-Computed Tomography
Photons
Gamma Cameras
Thorax
Research

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

Performance characteristics of transmission imaging using a uniform sheet source with parallel-hole collimation. / Cao, Zongjian; Tsui, Beniamin M W.

In: Medical Physics, Vol. 19, No. 5, 01.01.1992, p. 1205-1212.

Research output: Contribution to journalArticle

@article{5a6a42f1b4f6400db07c17e34ff07be5,
title = "Performance characteristics of transmission imaging using a uniform sheet source with parallel-hole collimation",
abstract = "Transmission imaging is receiving increasing attention in SPECT due to the need to compensate for nonuniform attenuation in cardiac-chest SPECT. The quality of a transmission image has an important effect on the measured attenuation distribution. To improve image quality, knowledge of the performance characteristics of a transmission imaging system is essential. The characteristics, spatial resolution, detection efficiency, photon flux, and exposure to the object, of a transmission imaging system consisting of a gamma camera and a uniform sheet source have been studied. The results demonstrate that spatial resolution of a transmission imaging system can be improved by use of a high-resolution source collimator at the price of a moderate decrease in detection efficiency, in comparison to the uncollimated case. Also, the source collimator significantly reduces the photon flux and exposure to the object. This investigation suggests that a high-resolution collimator be used with an intense sheet source to improve spatial resolution and reduce statistical noise with low exposure to the patient. This research further suggests that the amount of source activity is determined by the requirement of image quality, detection geometry, and allowed absorbed dose to the patient.",
author = "Zongjian Cao and Tsui, {Beniamin M W}",
year = "1992",
month = "1",
day = "1",
doi = "10.1118/1.596752",
language = "English (US)",
volume = "19",
pages = "1205--1212",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "5",

}

TY - JOUR

T1 - Performance characteristics of transmission imaging using a uniform sheet source with parallel-hole collimation

AU - Cao, Zongjian

AU - Tsui, Beniamin M W

PY - 1992/1/1

Y1 - 1992/1/1

N2 - Transmission imaging is receiving increasing attention in SPECT due to the need to compensate for nonuniform attenuation in cardiac-chest SPECT. The quality of a transmission image has an important effect on the measured attenuation distribution. To improve image quality, knowledge of the performance characteristics of a transmission imaging system is essential. The characteristics, spatial resolution, detection efficiency, photon flux, and exposure to the object, of a transmission imaging system consisting of a gamma camera and a uniform sheet source have been studied. The results demonstrate that spatial resolution of a transmission imaging system can be improved by use of a high-resolution source collimator at the price of a moderate decrease in detection efficiency, in comparison to the uncollimated case. Also, the source collimator significantly reduces the photon flux and exposure to the object. This investigation suggests that a high-resolution collimator be used with an intense sheet source to improve spatial resolution and reduce statistical noise with low exposure to the patient. This research further suggests that the amount of source activity is determined by the requirement of image quality, detection geometry, and allowed absorbed dose to the patient.

AB - Transmission imaging is receiving increasing attention in SPECT due to the need to compensate for nonuniform attenuation in cardiac-chest SPECT. The quality of a transmission image has an important effect on the measured attenuation distribution. To improve image quality, knowledge of the performance characteristics of a transmission imaging system is essential. The characteristics, spatial resolution, detection efficiency, photon flux, and exposure to the object, of a transmission imaging system consisting of a gamma camera and a uniform sheet source have been studied. The results demonstrate that spatial resolution of a transmission imaging system can be improved by use of a high-resolution source collimator at the price of a moderate decrease in detection efficiency, in comparison to the uncollimated case. Also, the source collimator significantly reduces the photon flux and exposure to the object. This investigation suggests that a high-resolution collimator be used with an intense sheet source to improve spatial resolution and reduce statistical noise with low exposure to the patient. This research further suggests that the amount of source activity is determined by the requirement of image quality, detection geometry, and allowed absorbed dose to the patient.

UR - http://www.scopus.com/inward/record.url?scp=0026806520&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0026806520&partnerID=8YFLogxK

U2 - 10.1118/1.596752

DO - 10.1118/1.596752

M3 - Article

C2 - 1435600

AN - SCOPUS:0026806520

VL - 19

SP - 1205

EP - 1212

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 5

ER -