Phantom evaluation of simultaneous thallium-201/technetium-99m aquisition in single-photon emission tomography

Zongjian Cao, Charles C. Chen, Christophe Maunoury, Lawrence E. Holder, Titus C. Abraham, Ann Tehan

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

This study investigates downscatter effects in cardiac single-photon emission tomographic studies with simultaneous thallium-201/technetium-99m acquisition, and evaluated a previously proposed subtraction technique for downscatter compensation. Ten studies were carried out with different defect sizes and locations and varying activity distributions using four energy windows: 70±10% keV, 140±10% keV, 100±10% KeV, and 103±16% keV. The subtraction technique used the 100-or 103-keV data to remove scattered 99mTc counts from the 70-keV data. The size and contrast of infarcts in the dual-isotope 70-keV image were artificially decreased compared to those in the 140-keV image, caused by scattered 99mTe counts that were comparable to the primary 201T1 counts in the 70-keV window. The subtraction technique produced larger defects and more heterogeneous activity in the myocardial wall in dual-isotope 70-keV images compared to the corresponding 201T1-only images. These artifacts were caused by the markedly different spatial distributions of scattered 99mTe counts in the 100-keV (or 103-keV) window as compared with the 70-keV window. It is concluded that scattered 99mTc photons may cause overestimation of ischemia and myocardial viability in simultaneous dual-isotope patient studies. The proposed subtraction technique was inaccurate and produced image artifacts. Adequate downscatter compensation methods must be developed before applying simultaneous 201T1/99mTC acquisition in clinical practice.

Original languageEnglish (US)
Pages (from-to)1514-1520
Number of pages7
JournalEuropean Journal Of Nuclear Medicine
Volume23
Issue number11
DOIs
StatePublished - Nov 21 1996

Fingerprint

Subtraction Technique
Thallium
Technetium
Photons
Tomography
Isotopes
Artifacts
Myocardial Ischemia

Keywords

  • Reconstruction
  • Simultaneous dual-isotope acquisition
  • Single-photon emission tomography

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Phantom evaluation of simultaneous thallium-201/technetium-99m aquisition in single-photon emission tomography. / Cao, Zongjian; Chen, Charles C.; Maunoury, Christophe; Holder, Lawrence E.; Abraham, Titus C.; Tehan, Ann.

In: European Journal Of Nuclear Medicine, Vol. 23, No. 11, 21.11.1996, p. 1514-1520.

Research output: Contribution to journalArticle

Cao, Zongjian ; Chen, Charles C. ; Maunoury, Christophe ; Holder, Lawrence E. ; Abraham, Titus C. ; Tehan, Ann. / Phantom evaluation of simultaneous thallium-201/technetium-99m aquisition in single-photon emission tomography. In: European Journal Of Nuclear Medicine. 1996 ; Vol. 23, No. 11. pp. 1514-1520.
@article{30ca465702c14a39848c5e5a76ec40cc,
title = "Phantom evaluation of simultaneous thallium-201/technetium-99m aquisition in single-photon emission tomography",
abstract = "This study investigates downscatter effects in cardiac single-photon emission tomographic studies with simultaneous thallium-201/technetium-99m acquisition, and evaluated a previously proposed subtraction technique for downscatter compensation. Ten studies were carried out with different defect sizes and locations and varying activity distributions using four energy windows: 70±10{\%} keV, 140±10{\%} keV, 100±10{\%} KeV, and 103±16{\%} keV. The subtraction technique used the 100-or 103-keV data to remove scattered 99mTc counts from the 70-keV data. The size and contrast of infarcts in the dual-isotope 70-keV image were artificially decreased compared to those in the 140-keV image, caused by scattered 99mTe counts that were comparable to the primary 201T1 counts in the 70-keV window. The subtraction technique produced larger defects and more heterogeneous activity in the myocardial wall in dual-isotope 70-keV images compared to the corresponding 201T1-only images. These artifacts were caused by the markedly different spatial distributions of scattered 99mTe counts in the 100-keV (or 103-keV) window as compared with the 70-keV window. It is concluded that scattered 99mTc photons may cause overestimation of ischemia and myocardial viability in simultaneous dual-isotope patient studies. The proposed subtraction technique was inaccurate and produced image artifacts. Adequate downscatter compensation methods must be developed before applying simultaneous 201T1/99mTC acquisition in clinical practice.",
keywords = "Reconstruction, Simultaneous dual-isotope acquisition, Single-photon emission tomography",
author = "Zongjian Cao and Chen, {Charles C.} and Christophe Maunoury and Holder, {Lawrence E.} and Abraham, {Titus C.} and Ann Tehan",
year = "1996",
month = "11",
day = "21",
doi = "10.1007/BF01254477",
language = "English (US)",
volume = "23",
pages = "1514--1520",
journal = "European Journal of Nuclear Medicine and Molecular Imaging",
issn = "1619-7070",
publisher = "Springer Verlag",
number = "11",

}

TY - JOUR

T1 - Phantom evaluation of simultaneous thallium-201/technetium-99m aquisition in single-photon emission tomography

AU - Cao, Zongjian

AU - Chen, Charles C.

AU - Maunoury, Christophe

AU - Holder, Lawrence E.

AU - Abraham, Titus C.

AU - Tehan, Ann

PY - 1996/11/21

Y1 - 1996/11/21

N2 - This study investigates downscatter effects in cardiac single-photon emission tomographic studies with simultaneous thallium-201/technetium-99m acquisition, and evaluated a previously proposed subtraction technique for downscatter compensation. Ten studies were carried out with different defect sizes and locations and varying activity distributions using four energy windows: 70±10% keV, 140±10% keV, 100±10% KeV, and 103±16% keV. The subtraction technique used the 100-or 103-keV data to remove scattered 99mTc counts from the 70-keV data. The size and contrast of infarcts in the dual-isotope 70-keV image were artificially decreased compared to those in the 140-keV image, caused by scattered 99mTe counts that were comparable to the primary 201T1 counts in the 70-keV window. The subtraction technique produced larger defects and more heterogeneous activity in the myocardial wall in dual-isotope 70-keV images compared to the corresponding 201T1-only images. These artifacts were caused by the markedly different spatial distributions of scattered 99mTe counts in the 100-keV (or 103-keV) window as compared with the 70-keV window. It is concluded that scattered 99mTc photons may cause overestimation of ischemia and myocardial viability in simultaneous dual-isotope patient studies. The proposed subtraction technique was inaccurate and produced image artifacts. Adequate downscatter compensation methods must be developed before applying simultaneous 201T1/99mTC acquisition in clinical practice.

AB - This study investigates downscatter effects in cardiac single-photon emission tomographic studies with simultaneous thallium-201/technetium-99m acquisition, and evaluated a previously proposed subtraction technique for downscatter compensation. Ten studies were carried out with different defect sizes and locations and varying activity distributions using four energy windows: 70±10% keV, 140±10% keV, 100±10% KeV, and 103±16% keV. The subtraction technique used the 100-or 103-keV data to remove scattered 99mTc counts from the 70-keV data. The size and contrast of infarcts in the dual-isotope 70-keV image were artificially decreased compared to those in the 140-keV image, caused by scattered 99mTe counts that were comparable to the primary 201T1 counts in the 70-keV window. The subtraction technique produced larger defects and more heterogeneous activity in the myocardial wall in dual-isotope 70-keV images compared to the corresponding 201T1-only images. These artifacts were caused by the markedly different spatial distributions of scattered 99mTe counts in the 100-keV (or 103-keV) window as compared with the 70-keV window. It is concluded that scattered 99mTc photons may cause overestimation of ischemia and myocardial viability in simultaneous dual-isotope patient studies. The proposed subtraction technique was inaccurate and produced image artifacts. Adequate downscatter compensation methods must be developed before applying simultaneous 201T1/99mTC acquisition in clinical practice.

KW - Reconstruction

KW - Simultaneous dual-isotope acquisition

KW - Single-photon emission tomography

UR - http://www.scopus.com/inward/record.url?scp=0029916046&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029916046&partnerID=8YFLogxK

U2 - 10.1007/BF01254477

DO - 10.1007/BF01254477

M3 - Article

C2 - 8854851

AN - SCOPUS:0029916046

VL - 23

SP - 1514

EP - 1520

JO - European Journal of Nuclear Medicine and Molecular Imaging

JF - European Journal of Nuclear Medicine and Molecular Imaging

SN - 1619-7070

IS - 11

ER -